Levine P A, Moran M J
St. Jude Medical, CRMD, Sylmar, California 91342, USA.
Pacing Clin Electrophysiol. 2000 Oct;23(10 Pt 1):1570-2. doi: 10.1046/j.1460-9592.2000.01570.x.
The normal pacing system function associated with the AutoCapture Pacing System algorithm triggered concern on the part of the clinical staff caring for the patient when the initiation of a threshold search sequence was detected during a transtelephonic follow-up evaluation. The analysis of the rhythm demonstrates that the behavior of the system is normal and consistent with the design of the AutoCapture algorithm. As a variety of new algorithms are introduced, similar unexpected but normal behaviors can be anticipated.