Computer-assisted eye examinations IV. "Additive" lens systems in eye refractors.

作者信息

Lang M M, Marg E

出版信息

Am J Optom Physiol Opt. 1975 Aug;52(8):533-40. doi: 10.1097/00006324-197508000-00004.

Abstract

After a brief introduction to the optical principles and terms, formulas are derived which demonstrate that with conventional methods and designs it is not possible to summate with accuracy the labeled power of three or more ophthalmic lenses in tandem. The effective power of sphero-cylindrical combinations in all currently available eye refractors, when powers of eight to ten diopters are exceeded, may differ from the summated labeled power by a quarter diopter or more, which can be clinically significant. This error can be directly and readily eliminated only in computer-actuated refractors by including a computer program which calculates the effective power and makes the necessary correction.

摘要

在对光学原理和术语进行简要介绍之后,推导出了公式,这些公式表明,使用传统方法和设计,不可能精确地叠加三个或更多串联的眼科镜片的标称度数。在所有目前可用的验光仪中,当球柱镜组合的度数超过8至10屈光度时,其有效度数可能与叠加后的标称度数相差四分之一屈光度或更多,这在临床上可能具有重要意义。只有在计算机驱动的验光仪中,通过包含一个计算有效度数并进行必要校正的计算机程序,才能直接且容易地消除这种误差。