Suppr超能文献

巩膜接触镜的屈光度分布及体外光学质量:孔径和屈光度的影响。

Power Profiles and In Vitro Optical Quality of Scleral Contact Lenses: Effect of the Aperture and Power.

作者信息

Domínguez-Vicent Alberto, Esteve-Taboada Jose Juan, Recchioni Alberto, Brautaset Rune

机构信息

Optometry Research Group (GIO), Department of Optics (A.D.-V., J.J.E.-T., A.R.), University of Valencia, Burjasot, Spain; and Unit of Optometry (R.B.), Department of Clinical Neuroscience, Karolinska Institutet, Stockholm, Sweden.

出版信息

Eye Contact Lens. 2018 May;44(3):149-158. doi: 10.1097/ICL.0000000000000345.

Abstract

OBJECTIVE

To assess the power profile and in vitro optical quality of scleral contact lenses with different powers as a function of the optical aperture.

METHODS

The mini and semiscleral contact lenses (Procornea) were measured for five powers per design. The NIMO TR-1504 (Lambda-X) was used to assess the power profile and Zernike coefficients of each contact lens. Ten measurements per lens were taken at 3- and 6-mm apertures. Furthermore, the optical quality of each lens was described in Zernike coefficients, modulation transfer function, and point spread function (PSF). A convolution of each lens PSF with an eye-chart image was also computed.

RESULTS

The optical power fluctuated less than 0.5 diopters (D) along the optical zone of each lens. However, the optical power obtained for some lenses did not match with its corresponding nominal one, the maximum difference being 0.5 D. In optical quality, small differences were obtained among all lenses within the same design. Although significant differences were obtained among lenses (P<0.05), these showed small impact in the image quality of each convolution.

CONCLUSIONS

Insignificant power fluctuations were obtained along the optical zone measured for each scleral lens. Additionally, the optical quality of both lenses has showed to be independent of the lens power within the same aperture.

摘要

目的

评估不同屈光度的巩膜接触镜的屈光力分布及体外光学质量随光学孔径的变化情况。

方法

对每种设计的迷你型和半巩膜接触镜(Procornea)测量五种屈光度。使用NIMO TR - 1504(Lambda - X)评估每个接触镜的屈光力分布和泽尼克系数。在3毫米和6毫米孔径处对每个镜片进行十次测量。此外,用泽尼克系数、调制传递函数和点扩散函数(PSF)描述每个镜片的光学质量。还计算了每个镜片PSF与视力表图像的卷积。

结果

沿每个镜片的光学区,屈光力波动小于0.5屈光度(D)。然而,一些镜片测得的屈光力与其相应的标称值不匹配,最大差值为0.5 D。在光学质量方面,同一设计内的所有镜片之间存在微小差异。虽然镜片之间存在显著差异(P<0.05),但这些差异对每个卷积的图像质量影响较小。

结论

在测量的每个巩膜镜片的光学区内,屈光力波动不显著。此外,在相同孔径内,两种镜片的光学质量均显示与镜片屈光度无关。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验