Suppr超能文献

使用光学无框架计算机辅助指针系统在侧颅底进行最佳配准(匹配)的实践要点。

Practical aspects for optimal registration (matching) on the lateral skull base with an optical frameless computer-aided pointer system.

作者信息

Caversaccio M, Zulliger D, Bächler R, Nolte L P, Häusler R

机构信息

Department of Otorhinolaryngology, Head and Neck Surgery, Inselspital, University of Bern, Switzerland.

出版信息

Am J Otol. 2000 Nov;21(6):863-70.

Abstract

HYPOTHESIS

Paired-point matching and surface matching are highly accurate when used on the lateral skull base with an optical computer-aided surgery system.

BACKGROUND

Computer-aided surgery on the lateral skull base can be done with a pointer system or with the microscope. An optical pointer system that uses anatomic landmarks and surface points has been developed in Bern, Switzerland.

METHODS

Axial computed tomography of a cadaver skull was performed. The images were processed on a computer workstation. An infrared camera was used. A reference base mounted on the head and a needle pointer, both equipped with light-emitting diodes, were used. Different anatomic landmarks were determined on the computer image of the skull and were compared with the actual anatomic markers taken on the bone or on the skin simulation material, and the target error was defined. In a second step, additional surface points on different regions of the skull were taken, and the inaccuracy from the target was redetermined.

RESULTS

The authors found a mean average error in accuracy from the target with paired-point matching alone in the best series of 0.79 mm. Under skin simulation, they found a deterioration with paired-point matching alone but an improvement in accuracy with surface matching.

CONCLUSION

For this navigation system, it is recommended that the following five anatomic points be selected for matching of the lateral skull base: the tip of the mastoid, the mastoid foramen, the umbo, the frontozygomatic suture, and the anterior nasal spine. For additional accuracy in clinical situations, surface matching is recommended.

摘要

假设

在使用光学计算机辅助手术系统对侧颅底进行操作时,配对点匹配和表面匹配具有高度准确性。

背景

侧颅底的计算机辅助手术可以通过指针系统或显微镜来完成。瑞士伯尔尼已开发出一种利用解剖标志和表面点的光学指针系统。

方法

对一具尸体颅骨进行轴向计算机断层扫描。图像在计算机工作站上进行处理。使用了一台红外摄像机。使用了安装在头部的参考基座和配备发光二极管的针状指针。在颅骨的计算机图像上确定不同的解剖标志,并与在骨或皮肤模拟材料上获取的实际解剖标记进行比较,定义目标误差。第二步,在颅骨的不同区域获取额外的表面点,并重新确定与目标的误差。

结果

作者发现,在最佳系列中,仅使用配对点匹配时,与目标的平均误差为0.79毫米。在皮肤模拟下,他们发现仅使用配对点匹配时误差增大,但表面匹配时准确性提高。

结论

对于该导航系统,建议选择以下五个解剖点用于侧颅底匹配:乳突尖、乳突孔、鼓膜脐、额颧缝和前鼻棘。为在临床情况下获得更高的准确性,建议采用表面匹配。