Hampl J, Hall M, Mufti N A, Yao Y M, MacQueen D B, Wright W H, Cooper D E
SRI International, Menlo Park, California, 94025, USA.
Anal Biochem. 2001 Jan 15;288(2):176-87. doi: 10.1006/abio.2000.4902.
Immunochromatographic assays have become popular diagnostic tools in a variety of settings because they are sensitive, fast, and easy to use. Here, we describe the use of a novel reporter, upconverting phosphors (UCP), in this assay format. UCP are submicron-sized, inorganic crystals that are excited with infrared light and that emit photons in the visible range depending on the ion composition of the crystal. Using human chorionic gonadotropin (hCG) as a model analyte to describe the properties of phosphors in immunochromatographic assays, a detection limit of 10 pg hCG in a 100-microl sample has been achieved on a regular basis, with occasional detection of 1 pg hCG. This represents at least a 10-fold improvement over conventional reporter systems such as colloidal gold or colored latex beads. Quantitation of analytes is possible over at least 3 orders of magnitude. Furthermore, an example is given of how UCP can be used for analyte multiplexing using a two-plexed wick for the detection of mouse IgG and ovalbumin. Thus, UCP lateral flow assays can be used for applications that are currently limited by assay sensitivity, and they can increase the probability of a diagnosis by verifying the presence of several analytes in the same sample.
免疫层析测定法因其灵敏、快速且易于使用,已成为各种环境中常用的诊断工具。在此,我们描述了一种新型报告分子——上转换荧光粉(UCP)在这种测定形式中的应用。UCP是亚微米级的无机晶体,由红外光激发,并根据晶体的离子组成在可见光范围内发射光子。以人绒毛膜促性腺激素(hCG)作为模型分析物来描述免疫层析测定法中荧光粉的特性,在100微升样品中,hCG的检测限通常可达10皮克,偶尔能检测到1皮克。这比传统报告系统如胶体金或彩色乳胶珠至少提高了10倍。分析物的定量至少可在3个数量级范围内进行。此外,还给出了一个例子,说明如何使用UCP通过双重复用灯芯检测小鼠IgG和卵清蛋白来实现分析物的多重检测。因此,UCP侧向流动测定法可用于目前受测定灵敏度限制的应用,并可通过在同一样品中验证多种分析物的存在来提高诊断的可能性。