Investigation on the ability of an ultrasound bubble detector to deliver size measurements of gaseous bubbles in fluid lines by using a glass bead model.

作者信息

Eitschberger S, Henseler A, Krasenbrink B, Oedekoven B, Mottaghy K

机构信息

Institute for Physiology, University Hospital of the Technical University of Aachen, Germany.

出版信息

ASAIO J. 2001 Jan-Feb;47(1):18-24. doi: 10.1097/00002480-200101000-00006.

DOI:10.1097/00002480-200101000-00006
PMID:11199308
Abstract

Detectors based on ultrasonic principles are today's state of the art devices to detect gaseous bubbles that may be present in extracorporeal circuits (ECC) for various reasons. Referring to theoretical considerations and other studies, it also seems possible to use this technology to measure the size of detected bubbles, thus offering the chance to evaluate their potential hazardous effect if introduced into a patient's circulation. Based on these considerations, a commercially available ultrasound bubble detector has been developed by Hatteland Instrumentering, Norway, to deliver bubble size measurements by means of supplementary software. This device consists of an ultrasound sensor that can be clamped onto the ECC tubing, and the necessary electronic equipment to amplify and rectify the received signals. It is supplemented by software that processes these signals and presents them as specific data. On the basis of our knowledge and experience with bubble detection by ultrasound technology, we believe it is particularly difficult to meet all the requirements for size measurements, especially if these are to be achieved by using a mathematical procedure rather than exact devices. Therefore, we tried to evaluate the quality of the offered bubble detector in measuring bubble sizes. After establishing a standardized test stand, including a roller pump and a temperature sensor, we performed several sets of experiments using the manufacturers software and a program specifically designed at our department for this purpose. The first set revealed that the manufacturer's recommended calibration material did not meet essential requirements as established by other authors. Having solved that problem, we could actually demonstrate that the ultrasonic field, as generated by the bubble detector, has been correctly calculated by the manufacturer. Simply, it is a field having the strongest reflecting region in the center, subsequently losing strength toward the ECC tubing's edge. The following set of experiments revealed that the supplementary software not only does not compensate for the ultrasonic field's inhomogeneity, but, furthermore, delivers results that are inappropriate to the applied calibration material. In the last set of experiments, we were able to demonstrate that the signals as recorded by the bubble detector heavily depend upon the circulating fluid's temperature, a fact that the manufacturer does not address. Therefore, it seems impossible to resolve all these sensor related problems by ever-increasing mathematical intervention. We believe it is more appropriate to develop a new kind of ultrasound device, free of these shortcomings. This seems to be particularly useful, because the problem of determining the size of gaseous bubbles in ECC is not yet solved.

摘要

相似文献

1
Investigation on the ability of an ultrasound bubble detector to deliver size measurements of gaseous bubbles in fluid lines by using a glass bead model.
ASAIO J. 2001 Jan-Feb;47(1):18-24. doi: 10.1097/00002480-200101000-00006.
2
The effectiveness of low-prime cardiopulmonary bypass circuits at removing gaseous emboli.低预充量体外循环回路清除气态栓子的有效性。
J Extra Corpor Technol. 2004 Dec;36(4):336-42.
3
In vitro evaluation of the air separation ability of four cardiovascular manufacturer extracorporeal circuit designs.四种心血管设备制造商体外循环设计的空气分离能力的体外评估。
J Extra Corpor Technol. 2006 Sep;38(3):206-13.
4
Ultrasound destruction of air microemboli as a novel approach to brain protection in cardiac surgery.超声破坏空气微栓子:心脏手术中脑保护的新方法。
J Cardiothorac Vasc Anesth. 2013 Oct;27(5):876-83. doi: 10.1053/j.jvca.2013.01.020. Epub 2013 Jun 18.
5
Deairing of the venous drainage in standard extracorporeal circulation results in a profound reduction of arterial micro bubbles.标准体外循环中静脉引流去气可导致动脉微泡显著减少。
Thorac Cardiovasc Surg. 2006 Feb;54(1):39-41. doi: 10.1055/s-2005-865874.
6
A technique for automatic tubing occlusion in response to air bubble detection when using a centrifugal pump.
J Extra Corpor Technol. 1990;22(2):98-100.
7
[A simple test procedure for detection of micro-bubbles in extracorporeal blood circulation].
Biomed Tech (Berl). 1993 Mar;38(3):34-7. doi: 10.1515/bmte.1993.38.3.34.
8
Acoustical bubble trapper applied to hemodialysis.应用于血液透析的声学气泡捕集器。
Ultrasound Med Biol. 2008 Apr;34(4):681-4. doi: 10.1016/j.ultrasmedbio.2007.09.014. Epub 2007 Nov 8.
9
Resonance ultrasonic measurements of microscopic gas bubbles.微观气泡的共振超声测量
Aviat Space Environ Med. 1976 Jul;47(7):777-81.
10
In vitro studies of gas bubble formation by mechanical heart valves.机械心脏瓣膜形成气泡的体外研究。
J Heart Valve Dis. 1999 Mar;8(2):186-96.

引用本文的文献

1
Observation of microbubbles during standard dialysis treatments.标准透析治疗期间微泡的观察。
Clin Kidney J. 2015 Aug;8(4):400-4. doi: 10.1093/ckj/sfv051. Epub 2015 Jul 3.
2
Evidence-based used, yet still controversial: the arterial filter.循证使用,但仍存在争议:动脉滤器
J Extra Corpor Technol. 2012 Mar;44(1):P27-30.
3
Post-arterial filter gaseous microemboli activity of five integral cardiotomy reservoirs during venting: an in vitro study.通气期间五个整体式心脏切开储血器的动脉后滤器气态微栓子活性:一项体外研究。
J Extra Corpor Technol. 2009 Mar;41(1):20-7.
4
Evaluation of the Doppler technique for fat emboli detection in an experimental flow model.在实验血流模型中评估多普勒技术检测脂肪栓子的效果。
J Extra Corpor Technol. 2008 Sep;40(3):175-83.
5
In vitro evaluation of the air separation ability of four cardiovascular manufacturer extracorporeal circuit designs.四种心血管设备制造商体外循环设计的空气分离能力的体外评估。
J Extra Corpor Technol. 2006 Sep;38(3):206-13.