Fleming J S, Goatman K A, Julyan P J, Boivin C M, Wilson M J, Barber R W, Bird N J, Fryer T D
Department of Nuclear Medicine, Southampton General Hospital, Shirley, UK.
Nucl Med Commun. 2000 Dec;21(12):1095-102. doi: 10.1097/00006231-200012000-00002.
The development of coincidence positron imaging using a gamma camera should significantly increase PET utilization. This study has compared the performance of three such systems with each other and with an equivalent single photon emission computed tomography (SPECT) measurement. A Carlson phantom, which consisted of a 20 cm diameter cylinder containing areas of uniform activity, hot lesions, cold lesions and orthogonal alternate hot and cold lines, was filled with an activity of 18F chosen to try and best exploit each camera and imaged for a fixed time (27 min). The image quality of 9 mm thick slices in each section of the phantom was compared visually. Several image quality parameters were also compared including line source resolution and noise equivalent count rate. There were considerable differences in PET image quality between the three cameras but all were at least as good as the SPECT image. The full width at half maximum (FWHM) resolution of all systems was similar (approximately 4.5 mm) but the maximum noise equivalent count rates in a 20 cm cylinder were significantly different (6.3, 2.6 and 1.6 kcps) (where cps is counts per second) and correlated with the phantom image quality.
使用伽马相机的符合正电子成像技术的发展应能显著提高正电子发射断层扫描(PET)的利用率。本研究比较了三种此类系统之间的性能,以及它们与等效单光子发射计算机断层扫描(SPECT)测量的性能。卡尔森体模由一个直径20厘米的圆柱体组成,内部包含均匀放射性区域、热病变、冷病变以及正交交替的热线条和冷线条,填充有选择的18F放射性活度,旨在尽力充分利用每台相机,并在固定时间(27分钟)内进行成像。对体模各部分9毫米厚切片的图像质量进行了视觉比较。还比较了几个图像质量参数,包括线源分辨率和噪声等效计数率。这三台相机的PET图像质量存在相当大的差异,但都至少与SPECT图像一样好。所有系统的半高宽(FWHM)分辨率相似(约4.5毫米),但在一个20厘米圆柱体中的最大噪声等效计数率有显著差异(6.3、2.6和1.6千计数每秒)(其中cps是每秒计数),且与体模图像质量相关。