Adam L E, Karp J S, Daube-Witherspoon M E, Smith R J
Department of Radiology, University of Pennsylvania, Philadelphia, Pennsylvania 19104, USA.
J Nucl Med. 2001 Dec;42(12):1821-30.
A whole-body PET scanner, without interplane septa, has been designed to achieve high performance in clinical applications. The C-PET scanner, an advancement of the PENN PET scanners, is unique in the use of 6 curved NaI(Tl) detectors (2.54 cm thick). The scanner has a ring diameter of 90 cm, a patient port diameter of 56 cm, and an axial field of view of 25.6 cm. A (137)Cs point source is used for transmission scans.
Following the protocols of the International Electrotechnical Commission ([IEC] 61675-1) and the National Electrical Manufacturers Association ([NEMA] NU-2-1994 and an updated version, NU2-2001), point and line sources, as well as uniform cylinders, were used to determine the performance characteristics of the C-PET scanner. An image-quality phantom and patient data were used to evaluate image quality under clinical scanning conditions. Data were rebinned with Fourier rebinning into 2-dimensional (slice-oriented) datasets and reconstructed with an iterative reconstruction algorithm.
The spatial resolution for a point source in the transaxial direction was 4.6 mm (full width at half maximum) at the center, and the axial resolution was 5.7 mm. For the NU2-1994 analysis, the sensitivity was 12.7 cps/Bq/mL (444 kcps/microCi/mL), the scatter fraction was 25%, and the peak noise equivalent count rate (NEC) for a uniform cylinder (diameter = 20 cm, length = 19 cm) was 49 kcps at an activity concentration of 11.2 kBq/mL. For the IEC protocol, the peak NEC was 41 kcps at 12.3 kBq/mL, and for the NU2-2001 protocol, the peak NEC was 14 kcps at 3.8 kBq/mL. The NU2-2001 NEC value differed significantly because of differences in the data analysis and the use of a 70-cm-long phantom.
Compared with previous PENN PET scanners, the C-PET, with its curved detectors and improvements in pulse shaping, integration dead time, and triggering, has an improved count-rate capability and spatial resolution. With the refinements in the singles transmission technique and iterative reconstruction, image quality is improved and scan time is shortened. With single-event transmission scans interleaved between sequential emission scans, a whole-body study can be completed in <1 h. Overall, C-PET is a cost-effective PET scanner that performs well in a broad variety of clinical applications.
一种无层间隔板的全身正电子发射断层扫描仪(PET)已被设计用于在临床应用中实现高性能。C-PET扫描仪是宾夕法尼亚PET扫描仪的升级版,其独特之处在于使用了6个弯曲的碘化钠(铊)探测器(厚2.54厘米)。该扫描仪的环直径为90厘米,患者端口直径为56厘米,轴向视野为25.6厘米。使用一个铯-137点源进行透射扫描。
按照国际电工委员会(IEC 61675-1)和美国国家电气制造商协会(NEMA NU-2-1994及其更新版本NU2-2001)的协议,使用点源、线源以及均匀圆柱体来确定C-PET扫描仪的性能特征。使用图像质量体模和患者数据来评估临床扫描条件下的图像质量。数据通过傅里叶重排重新组合为二维(面向切片)数据集,并使用迭代重建算法进行重建。
在横向方向上,点源在中心处的空间分辨率为4.6毫米(半高宽),轴向分辨率为5.7毫米。对于NU2-1994分析,灵敏度为12.7计数每秒/贝克勒尔/毫升(444千计数每秒/微居里/毫升),散射分数为25%,对于一个均匀圆柱体(直径 = 20厘米,长度 = 19厘米),在活度浓度为11.2千贝克勒尔/毫升时,峰值噪声等效计数率(NEC)为49千计数每秒。对于IEC协议,在12.3千贝克勒尔/毫升时峰值NEC为41千计数每秒,对于NU2-2001协议,在3.8千贝克勒尔/毫升时峰值NEC为14千计数每秒。由于数据分析和使用70厘米长体模的差异,NU2-2001的NEC值有显著不同。
与之前的宾夕法尼亚PET扫描仪相比,C-PET凭借其弯曲探测器以及在脉冲整形、积分死时间和触发方面的改进,具有更高的计数率能力和空间分辨率。随着单光子透射技术和迭代重建的改进,图像质量得到提高且扫描时间缩短。通过在连续发射扫描之间穿插单事件透射扫描,全身检查可在1小时内完成。总体而言,C-PET是一种性价比高的PET扫描仪,在多种临床应用中表现良好。