ECAT ART - a continuously rotating PET camera: performance characteristics, initial clinical studies, and installation considerations in a nuclear medicine department.

作者信息

Bailey D L, Young H, Bloomfield P M, Meikle S R, Glass D, Myers M J, Spinks T J, Watson C C, Luk P, Peters A M, Jones T

机构信息

MRC Cyclotron Unit, Hammersmith Hospital, London, UK.

出版信息

Eur J Nucl Med. 1997 Jan;24(1):6-15. doi: 10.1007/BF01728302.

Abstract

Advances in fully three-dimensional (3D) image reconstruction techniques have permitted the development of a commercial, rotating, partial ring, fully 3D positron emission tomographic (PET) scanner, the ECAT ART. The system has less than one-half the number of bismuth germanate detectors compared with a full ring scanner with the equivalent field of view, resulting in reduced capital cost. The performance characteristics, implications for installation in a nuclear medicine department, and clinical utility of the scanner are presented in this report. The sensitivity (20 cm diameterx20 cm long cylindrical phantom, no scatter correction) is 11400 cps.kBq-1.ml-1. This compares with 5800 and 40500 cps.kBq-1.ml-1 in 2D and 3D respectively for the equivalent full ring scanner (ECAT EXACT). With an energy window of 350-650 keV the maximum noise equivalent count (NEC) rate was 27 kcps at a radioactivity concentration of approximately 15 kBq.ml-1 in the cylinder. Spatial resolution is approximately 6 mm full width at half maximum on axis degrading to just under 8 mm at a distance of 20 cm off axis. Installation and use within the nuclear medicine department does not appreciably increase background levels of radiation on gamma cameras in adjacent rooms and the dose rate to an operator in the same room is 2 microSv. h-1 for a typical fluorine-18 fluorodeoxyglucose (18F-FDG) study with an initial injected activity of 370 MBq. The scanner has been used for clinical imaging with18F-FDG for neurological and oncological applications. Its novel use for imaging iron-52 transferrin for localising erythropoietic activity demonstrates its sensitivity and resolution advantages over a conventional dual-headed gamma camera. The ECAT ART provides a viable alternative to conventional full ring PET scanners without compromising the performance required for clinical PET imaging.

摘要

全三维(3D)图像重建技术的进步推动了一款商用旋转部分环全3D正电子发射断层扫描(PET)扫描仪ECAT ART的开发。与具有同等视野的全环扫描仪相比,该系统的锗酸铋探测器数量减少了一半以上,从而降低了资金成本。本报告介绍了该扫描仪的性能特点、在核医学科安装的影响以及临床实用性。灵敏度(20厘米直径×20厘米长的圆柱形体模,无散射校正)为11400 cps.kBq-1.ml-1。相比之下,同等全环扫描仪(ECAT EXACT)在二维和三维模式下的灵敏度分别为5800和40500 cps.kBq-1.ml-1。在350 - 650 keV的能量窗口下,圆柱体中放射性浓度约为15 kBq.ml-1时,最大噪声等效计数(NEC)率为27 kcps。空间分辨率在轴向上约为半高宽6毫米,在离轴20厘米处降至略低于8毫米。在核医学科安装和使用该扫描仪不会显著增加相邻房间γ相机的背景辐射水平,对于典型的初始注射活度为370 MBq的氟-18氟脱氧葡萄糖(18F-FDG)研究,同一房间内操作人员的剂量率为2微Sv.h-1。该扫描仪已用于18F-FDG的神经学和肿瘤学临床成像。其在成像铁-52转铁蛋白以定位造血活性方面的新用途证明了其相对于传统双头γ相机的灵敏度和分辨率优势。ECAT ART为传统全环PET扫描仪提供了一种可行的替代方案,同时不影响临床PET成像所需的性能。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验