Hausleiter J, Li A, Makkar R, Berman D, Robinson A, Litvack F, Eigler N, Whiting J
Department of Medical Physics and Imaging, Cedars-Sinai Medical Center, Davis Building #6065, 8700 Beverly Boulevard, Los Angeles, CA 90048, USA.
Cardiovasc Radiat Med. 2000 Jan-Mar;2(1):7-10.
We are reporting the first case of an accidental radioactive 188Re leakage of a liquid-filled balloon system. Different analytical methods estimated that approximately 4 mCi 188Re were released. The radiation burden was reduced considerably by the combined therapy with perchlorate and forced volume diuresis. Estimated exposures to all organs were very low with 1.8 rad. A total body nuclear scintigraphy demonstrated uniform 188Re distribution, without specific organ concentration.
我们报告首例充液气球系统意外放射性188铼泄漏事件。不同分析方法估计约4毫居里的188铼被释放。通过高氯酸盐联合治疗和强制大量利尿,辐射负担显著降低。所有器官的估计暴露量非常低,为1.8拉德。全身核闪烁显像显示188铼分布均匀,无特定器官浓聚。