A tandem column ensemble with an atmospheric pressure junction-point vent for high-speed GC with selective control of peak-pair separation.

作者信息

Veriotti T, Sacks R

机构信息

Department of Chemistry, University of Michigan, Ann Arbor 48109, USA.

出版信息

Anal Chem. 2001 Feb 15;73(4):813-9. doi: 10.1021/ac001028w.

Abstract

A series-coupled (tandem) ensemble of two capillary GC columns using different stationary phases and a pneumatically actuated low-volume valve connecting the column junction point to an atmospheric-pressure vent line is used to adjust the ensemble separation of selected pairs of target compounds. The valve is normally closed, and the pressure at the column junction point assumes the value that would occur in the absence of any other connections. The valve can be opened for brief periods of time, thus producing pulses of atmospheric pressure at the column junction point. If a component pair is separated by the first column but coelutes from the column ensemble, the ensemble separation can be increased if a pulse occurs when one of the components has migrated across the column junction but the second component is still on the first column. All of the mixture components that are on the same column during the time that the valve is open (pulse duration) will be shifted to either larger or smaller retention times, but the pattern of peaks (elution order) for these components from the column ensemble will be relatively unaffected by the pressure pulse. Multiple pulses can be used to enhance the separation of different component pairs, which sequentially reach the column junction point. Performance of the valve-operated system is described. Time-of-flight mass spectrometry with time-array detection is used to examine the effects of pulse duration on the separation achieved for different component pairs.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验