al-Hazaimeh N, Gutteridge D L
Division of Restorative Dentistry, Leeds Dental Institute, Clarendon Way, Leeds, LS2 9LU, UK.
Int Endod J. 2001 Jan;34(1):40-6. doi: 10.1046/j.1365-2591.2001.00351.x.
This in vitro study investigated the effect of a ferrule preparation on the fracture resistance of crowned central incisors incorporating a prefabricated post (Parapost) cemented with Panavia-Ex and with a composite core.
The test group consisted of 10 post crowned natural central incisor teeth with a 2-mm wide ferrule preparation, whilst the control group of 10 teeth had no ferrule. The specimens were mounted on a Lloyd universal testing machine and a compressive load was applied at an angle of 135 degrees to the palatal surface of the crown until failure occurred.
In both groups, failure occurred at higher loads compared with previous studies. The mean failure load was 1218 N for the unferruled group and 1407 N for the ferruled group. Statistical analysis showed no significant difference between the groups.
Under the conditions of this study, when composite cement and core materials are utilized with a Parapost prefabricated system in vitro the additional use of a ferrule preparation has no benefit in terms of resistance to fracture.
本体外研究调查了套圈预备对采用帕拉派斯特(Parapost)预成桩并用帕纳维亚-Ex(Panavia-Ex)粘结且带有复合树脂核的冠修复中切牙抗折性能的影响。
试验组由10颗制备了2毫米宽套圈的桩冠修复天然中切牙组成,而对照组的10颗牙未制备套圈。将标本安装在劳埃德万能试验机上,并以135度角向牙冠腭面施加压缩载荷,直至发生破坏。
与以往研究相比,两组在更高载荷下发生破坏。无套圈组的平均破坏载荷为1218牛,有套圈组为1407牛。统计分析表明两组之间无显著差异。
在本研究条件下,体外使用复合树脂粘结剂和核材料以及帕拉派斯特预成系统时,额外制备套圈对提高抗折性能并无益处。