A comparison of canine and human UHMWPE acetabular component wear.

作者信息

Skurla C T, James S P

机构信息

Department of Mechanical Engineering, Colorado State University, Fort Collins, CO 80523-1374, USA.

出版信息

Biomed Sci Instrum. 2001;37:245-50.

PMID:11347396
Abstract

The objective of this study is to compare the wear patterns and wear rates of ultra high molecular weight polyethylene in acetabular components used in human and canine total hip arthroplasty. This comparison may help elucidate both the appropriateness of the dog as an orthopaedic model for humans as well as the clinical performance of canine total hip arthroplasties. Twenty-nine postmortem retrieved and three revision retrieved canine acetabular components were analyzed. The volumetric wear rates in the canine postmortem retrievals of 6.9 mm3/year for modular (BioMedtrix) implants and 8.3 mm3/yr for monolithic (Richards) implants were considerably lower than those of 35.0 mm3/yr measured in human postmortem retrievals, reported using a similar measurement technique. The canine revision retrieval volumetric wear rate of 34.8 mm3/yr was significantly higher than postmortem rates, yet considerably lower than the reported retrieval rates of 62.0 mm3/yr seen in humans. These results may have negative implications for the use of the dog as an orthopaedic model for humans in total hip arthroplasty studies.

摘要

相似文献

1
A comparison of canine and human UHMWPE acetabular component wear.
Biomed Sci Instrum. 2001;37:245-50.
2
Comparison of the in vivo wear rates of 43 surgically retrieved direct compression molded and ram extruded ultra high molecular weight polyethylene acetabular components.43个手术取出的直接模压成型和冲压挤出超高分子量聚乙烯髋臼部件的体内磨损率比较。
Biomed Sci Instrum. 1996;32:135-41.
3
Implant wear mechanisms--basic approach.植入物磨损机制——基本方法。
Biomed Mater. 2008 Dec;3(4):042001. doi: 10.1088/1748-6041/3/4/042001. Epub 2008 Sep 25.
4
Polyethylene sterilization and production affects wear in total hip arthroplasties.聚乙烯的灭菌和生产会影响全髋关节置换术的磨损情况。
Clin Orthop Relat Res. 2006 Dec;453:305-8. doi: 10.1097/01.blo.0000229348.10458.79.
5
[Polyethylene wear in periprosthetic tissue of failed cemented total hip arthroplasties].[骨水泥型全髋关节置换失败病例假体周围组织中的聚乙烯磨损]
Zhonghua Wai Ke Za Zhi. 1997 Aug;35(8):459-61.
6
[Measurement of acetabular polyethylene wear of total hip replacement, using a universal measuring microscope. Characteristics of measurements].[使用万能测量显微镜测量全髋关节置换术中髋臼聚乙烯磨损情况。测量特征]
Acta Chir Orthop Traumatol Cech. 2006;73(1):28-33.
7
In vivo oxidation of retrieved cross-linked ultra-high-molecular-weight polyethylene acetabular components with residual free radicals.体内回收的带有残留自由基的交联超高分子量聚乙烯髋臼部件的氧化
J Arthroplasty. 2006 Oct;21(7):1005-11. doi: 10.1016/j.arth.2005.07.019. Epub 2006 Jul 17.
8
[Method for assessment of distribution of UHMWPE wear particles in periprosthetic tissues in total hip arthroplasty].[全髋关节置换术中评估超高分子量聚乙烯磨损颗粒在假体周围组织中分布的方法]
Acta Chir Orthop Traumatol Cech. 2006 Aug;73(4):243-50.
9
Role of cyclic plastic deformation in the wear of UHMWPE acetabular cups.循环塑性变形在超高分子量聚乙烯髋臼杯磨损中的作用。
J Biomed Mater Res. 1995 May;29(5):619-26. doi: 10.1002/jbm.820290509.
10
Analysis of a retrieved UHMWPE acetabular cup crosslinked in air with 1000 kGy of gamma radiation.对一个在空气中用1000千戈瑞伽马射线交联的超高分子量聚乙烯髋臼杯的分析。
J Orthop Res. 2004 Jul;22(4):828-31. doi: 10.1016/j.orthres.2003.12.001.