The changes in delivered oxygen fractions using laerdal resuscitator bag with different types of reservoir.

作者信息

Nam S H, Kim K J, Nam Y T, Shim J K

机构信息

Department of Anesthesiology, Yonsei University College of Medicine, Seoul, Korea.

出版信息

Yonsei Med J. 2001 Apr;42(2):242-6. doi: 10.3349/ymj.2001.42.2.242.

Abstract

One of the disadvantages of the Laerdal resuscitator bag is that it does not deliver a high concentration of oxygen without a reservoir and an appropriate technique of ventilation. With a specific device that is able to compress a resuscitator bag mechanically at a regular volume, ventilator rate, and speed, we evaluated the effects of various factors (the tidal volume, the ventilator rate, the oxygen flow rate, the type of reservoir) of the Laerdal resuscitator bag during positive pressure ventilation that affect the delivered oxygen fraction (FDO2) and also whether 250 mL and 500 mL corrugated tubes could be used as substitutes for the reservoir bag. The 250 mL corrugated tube increased the FDO2 to over 96% with an oxygen flow rate of 15 L/min. The 500 mL corrugated tube increased the FDO2 to over 96% with an oxygen flow rate of 10 L/min regardless of the ventilator rate at a fixed tidal volume of 500 mL. At the identical fixed tidal volume of 500 mL, the 1,600 mL reservoir bag increased the FDO2 to over 92% with an oxygen flow rate of 5 L/min and to over 96% at 7.5 L/min regardless of the ventilator rate. We concluded that the FDO2 of the Laerdal resuscitator bag depends on various factors such as tidal volume, ventilator rate, oxygen flow rate, and type of reservoir and both the 250 mL and 500 mL corrugated tubes can be used as substitutes.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验