Tcholakian R K, Raad I I
Department of Interdisciplinary Research, University of Texas-Houston Medical School, University of Texas M. D. Anderson Cancer Center, Houston, Texas, USA.
Antimicrob Agents Chemother. 2001 Jul;45(7):1990-3. doi: 10.1128/AAC.45.7.1990-1993.2001.
This study was performed to test the long-term antimicrobial efficacy of impregnated silicone catheters comprising an antimicrobial layer sandwiched between an external surface sheath and a luminal surface silicone sheath. The design of the catheter permits the introduction of various antimicrobials in addition to anticoagulants or antifibrins in the antimicrobial layer and allows their gradual release over a period of months after insertion. The in vitro data presented show that the catheter can provide antimicrobial activity for 90 days, after being replated for 15 7-day cycles of replating. When the catheters were immersed in human serum and incubated at 37 degrees C, they demonstrated significant antimicrobial activity after more than 325 days of incubation. The significant long-term in vitro antimicrobial activity observed may imply effective in vivo activity for almost 1 year after insertion and could serve as a cost-effective alternative to surgically implantable silicone catheters.
本研究旨在测试一种浸渍硅胶导管的长期抗菌效果,该导管包含夹在外表面护套和管腔表面硅胶护套之间的抗菌层。导管的设计允许在抗菌层中除了添加抗凝剂或抗纤维蛋白外,还可引入各种抗菌剂,并使其在插入后的数月内逐渐释放。所呈现的体外数据表明,在进行15个7天的再接种循环后,该导管可提供90天的抗菌活性。当导管浸入人血清并在37℃下孵育时,孵育超过325天后它们仍表现出显著的抗菌活性。观察到的显著长期体外抗菌活性可能意味着插入后近1年内具有有效的体内活性,并且可以作为手术植入硅胶导管的一种经济有效的替代方案。