A novel method for continuous online glucose monitoring in humans: the comparative microdialysis technique.

作者信息

Hoss U, Kalatz B, Gessler R, Pfleiderer H J, Andreis E, Rutschmann M, Rinne H, Schoemaker M, Haug C, Fussgaenger R D

机构信息

Institute of Diabetes Technology at the University of Ulm, Germany.

出版信息

Diabetes Technol Ther. 2001 Summer;3(2):237-43. doi: 10.1089/152091501300209615.

Abstract

The aim of this study was to prove the feasibility of continuous subcutaneous glucose monitoring in humans using the comparative microdialysis technique (CMT). The performance of the CMT was determined by comparing tissue glucose values with venous or capillary blood glucose values in healthy volunteers and type 1 diabetic subjects. The CMT is a microdialysis-based system for continuous online glucose monitoring in humans. This technique does not require calibration by the patient. Physiological saline with glucose (5.5 mM) is pumped in a stop-flow mode through a microdialysis probe inserted into the abdominal s.c. tissue. Tissue glucose concentration is calculated by comparing the dialysate and perfusate glucose concentrations. The time delay due to the measurement process is 9 min. We tested the CMT on six healthy volunteers and six type 1 diabetic patients for 24 h in our clinical setting. Comparisons were made to HemoCue analyzer (Angelholm, Sweden) capillary blood glucose measurements (healthy volunteers) and to venous blood glucose concentration determined with a Hitachi analyzer (diabetic patients). The mean absolute relative error of the CMT glucose values from the blood glucose values was 17.8+/-15.5% (n = 167) for the healthy volunteers and 11.0+/-10.8% (n = 425) for the diabetic patients. The mean difference was 0.42+/-1.06 mM (healthy volunteers) and -0.17+/-1.22 mM (diabetic patients). Error grid analysis for the values obtained in diabetic patients demonstrated that 99% of CMT glucose values were within clinically acceptable regions (regions A and B of the Clarke Error Grid). The study results show that the CMT is an accurate technique for continuous online glucose monitoring.

摘要

本研究的目的是通过比较微透析技术(CMT)来证明连续皮下葡萄糖监测在人体中的可行性。通过比较健康志愿者和1型糖尿病患者的组织葡萄糖值与静脉或毛细血管血糖值来确定CMT的性能。CMT是一种基于微透析的人体连续在线葡萄糖监测系统。该技术无需患者进行校准。含葡萄糖(5.5 mM)的生理盐水以停流模式通过插入腹部皮下组织的微透析探头泵入。通过比较透析液和灌注液中的葡萄糖浓度来计算组织葡萄糖浓度。测量过程导致的时间延迟为9分钟。我们在临床环境中对6名健康志愿者和6名1型糖尿病患者进行了24小时的CMT测试。将结果与HemoCue分析仪(瑞典安吉尔霍尔姆)测量的毛细血管血糖值(健康志愿者)以及日立分析仪测定的静脉血糖浓度(糖尿病患者)进行比较。健康志愿者CMT葡萄糖值与血糖值的平均绝对相对误差为17.8±15.5%(n = 167),糖尿病患者为11.0±10.8%(n = 425)。平均差值在健康志愿者中为0.42±1.06 mM,在糖尿病患者中为-0.17±1.22 mM。对糖尿病患者获得的值进行误差网格分析表明,99%的CMT葡萄糖值在临床可接受区域内(克拉克误差网格的A区和B区)。研究结果表明,CMT是一种用于连续在线葡萄糖监测的准确技术。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验