Soares C G, Vynckier S, Järvinen H, Cross W G, Sipilä P, Flühs D, Schaeken B, Mourtada F A, Bass G A, Williams T T
National Institute of Standards and Technology, Gaithersburg, Maryland 20899, USA.
Med Phys. 2001 Jul;28(7):1373-84. doi: 10.1118/1.1376441.
An international intercomparison of the dosimetry of three beta particle emitting ophthalmic applicators was performed, which involved measurements with radiochromic film, thermoluminescence dosimeters (TLDs), alanine pellets, plastic scintillators, extrapolation ionization chambers, a small fixed-volume ionization chambers, a diode detector and a diamond detector. The sources studied were planar applicators of 90Sr-90Y and 106Ru-106Rh, and a concave applicator of 106Ru-106Rh. Comparisons were made of absolute dosimetry determined at 1 mm from the source surface in water or water-equivalent plastic, and relative dosimetry along and perpendicular to the source axes. The results of the intercomparison indicate that the various methods yield consistent absolute dosimetry results at the level of 10%-14% (one standard deviation) depending on the source. For relative dosimetry along the source axis at depths of 5 mm or less, the agreement was 3%-9% (one standard deviation) depending on the source and the depth. Crucial to the proper interpretation of the measurement results is an accurate knowledge of the detector geometry, i.e., sensitive volume and amount of insensitive covering material. From the results of these measurements, functions which describe the relative dose rate along and perpendicular to the source axes are suggested.
对三种发射β粒子的眼科敷贴器进行了剂量测定的国际比对,其中包括使用放射变色胶片、热释光剂量计(TLD)、丙氨酸颗粒、塑料闪烁体、外推电离室、小型固定体积电离室、二极管探测器和金刚石探测器进行测量。所研究的源为90Sr-90Y和106Ru-106Rh的平面敷贴器,以及106Ru-106Rh的凹面敷贴器。对在水或水等效塑料中距源表面1mm处测定的绝对剂量学,以及沿源轴和垂直于源轴的相对剂量学进行了比较。比对结果表明,根据源的不同,各种方法得出的绝对剂量学结果在10%-14%(一个标准差)的水平上是一致的。对于在5mm或更小深度沿源轴的相对剂量学,根据源和深度的不同,一致性为3%-9%(一个标准差)。准确了解探测器的几何形状,即灵敏体积和非灵敏覆盖材料的量,对于正确解释测量结果至关重要。根据这些测量结果,提出了描述沿源轴和垂直于源轴的相对剂量率的函数。