Extrapolation chamber measurements of 90Sr + 90Y beta-particle ophthalmic applicator dose rates.

作者信息

Deasy J O, Soares C G

机构信息

James Graham Brown Cancer Center, Department of Radiation Oncology, University of Louisville, Kentucky 40202.

出版信息

Med Phys. 1994 Jan;21(1):91-9. doi: 10.1118/1.597239.

Abstract

Aspects of extrapolation chamber dose-rate measurements of 90Sr + 90Y beta-particle ophthalmic applicators are examined in this report, including the proper choice of collector electrode size, the gap width over which the measurement should be done, the effect of the entrance window materials, and the stopping-power ratio. Experiments, a simple analytic model for the effect of chamber geometry and nonzero gap width, and more detailed Monte Carlo simulations were used. The variation of the planar flux density as a function of angle for a thick 90Sr + 90Y source was measured and used as input for the model. From Monte Carlo simulation, the dose rate for tissue irradiation falls off by 8% between the surface and a depth of 7 mg/cm2. The derivative of chamber ionization as a function of gap width, needed for the dose-rate calibration, increases rapidly as the gap width decreases, typically by a factor of about 2 between gap widths of 1.5 and 0.15 mm. About half of this change is due to ionizing electrons leaving the collection volume at the larger gap widths as shown by the analytic model; the rest of the change is due to ionizing electrons which backscatter from the collector electrode and its backing as shown by Monte Carlo simulations. The backscattering effect increases the derived surface dose by a factor of 1.46. A satisfactory dose-rate extrapolation is obtained from gap widths of 0.1-0.25 mm, where the total ionization current is observed to be nearly linear in gap width.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验