Extrapolation chamber measurements of 90Sr + 90Y beta-particle ophthalmic applicator dose rates.

作者信息

Deasy J O, Soares C G

机构信息

James Graham Brown Cancer Center, Department of Radiation Oncology, University of Louisville, Kentucky 40202.

出版信息

Med Phys. 1994 Jan;21(1):91-9. doi: 10.1118/1.597239.

Abstract

Aspects of extrapolation chamber dose-rate measurements of 90Sr + 90Y beta-particle ophthalmic applicators are examined in this report, including the proper choice of collector electrode size, the gap width over which the measurement should be done, the effect of the entrance window materials, and the stopping-power ratio. Experiments, a simple analytic model for the effect of chamber geometry and nonzero gap width, and more detailed Monte Carlo simulations were used. The variation of the planar flux density as a function of angle for a thick 90Sr + 90Y source was measured and used as input for the model. From Monte Carlo simulation, the dose rate for tissue irradiation falls off by 8% between the surface and a depth of 7 mg/cm2. The derivative of chamber ionization as a function of gap width, needed for the dose-rate calibration, increases rapidly as the gap width decreases, typically by a factor of about 2 between gap widths of 1.5 and 0.15 mm. About half of this change is due to ionizing electrons leaving the collection volume at the larger gap widths as shown by the analytic model; the rest of the change is due to ionizing electrons which backscatter from the collector electrode and its backing as shown by Monte Carlo simulations. The backscattering effect increases the derived surface dose by a factor of 1.46. A satisfactory dose-rate extrapolation is obtained from gap widths of 0.1-0.25 mm, where the total ionization current is observed to be nearly linear in gap width.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验