Dosimetry close to an 192Ir HDR source using N-vinylpyrrolidone based polymer gels and magnetic resonance imaging.

作者信息

Papagiannis P, Pappas E, Kipouros P, Angelopoulos A, Sakelliou L, Baras P, Karaiskos P, Seimenis I, Sandilos P, Baltas D

机构信息

Department of Physics, University of Athens, Greece.

出版信息

Med Phys. 2001 Jul;28(7):1416-26. doi: 10.1118/1.1382603.

Abstract

In this work, the utilization of polymer gel-MRI dosimetry for measurements at distances relevant to clinical brachytherapy and intravascular applications [i.e., in the mm range, where steep three-dimensional (3-D) dose gradients exist] is investigated using N-vinylpyrrolidone-based gels. Transverse axis radial dose distributions, dose distributions parallel to the source axis, and 2-D dose distributions around the commonly used microSelectron 192Ir HDR source are measured for single source dwell position irradiations. Experimental results are found in good agreement with verified Monte Carlo calculations, even for distances less than 3 mm from the source. The effect of various MRI parameters, such as slice thickness, slice mispositioning, and in-plane resolution, on the accuracy of the method is also investigated. Possible limitations of the method are discussed, and its' overall potential in brachytherapy dosimetry is evaluated. Experimental 2-D dose distributions for an intravascular application following the Paris irradiation protocol are compared to corresponding commercial treatment planning system calculations. Results suggest that polymer gel-MRI dosimetry is capable of experimentally verifying dose distributions in relevant clinical intravascular applications.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验