Krupp E M, Pécheyran C, Meffan-Main S, Donard O F
LCABIE/CNRS, Helioparc Pau Pyreneés, France.
Fresenius J Anal Chem. 2001 Jul;370(5):573-80. doi: 10.1007/s002160100778.
The precision and accuracy of lead isotope-ratio determination on a short transient signal has been assessed by coupling capillary gas chromatography to the Isoprobe (Micromass, UK), a single-focusing inductively coupled plasma mass spectrometer with multicollector detection. A T-piece connecting the GC transfer line to the torch enabled continuous aspiration of thallium solution for mass-bias correction. The volatile lead species PbEt4 was derivatized from NIST isotopic certified lead standard SRM 981 and different amounts of PbEt4 dissolved in iso-octane were injected into the GC. Chromatograms were recorded in multicollection mode by use of Faraday cups; seven isotopes (204Pb, 206Pb, 207Pb, 208Pb,