Krupp Eva M, Pécheyran Christophe, Meffan-Main Simon, Donard Olivier F X
Helioparc Pau Pyreneés, LCABIE/CNRS, 2 Avenue du President Angot, 64053 Pau, Cedex 9, France.
Anal Bioanal Chem. 2004 Jan;378(2):250-5. doi: 10.1007/s00216-003-2328-9. Epub 2003 Nov 14.
Capillary gas chromatography coupled to an inductively coupled plasma mass spectrometer with multiple-collector detection (GC-ICP-MCMS) has been used to assess the precision and instrumental mass bias in sulfur isotope-ratio determination for the gaseous sulfur species SF6. The isotopic composition of the compound was certified by the institute for reference materials and measurements (IRMM, Belgium) and is available as PIGS 2010. Integration of the peaks (peak half-width 1.4 s) was performed using a special peak-integration method based on definition of the integration area by assessment of a uniform isotope-ratio area within the chromatographic peak. Instrumental mass bias was determined to be approximately 12% per mass unit and proved to be stable in the concentration range measured. Replicate injections of 2, 10, 20, and 30 ng (as S) SF6 diluted in argon gave precision for the 32S/34S ratio from 0.6% RSD for 2-ng injections to 0.03% RSD for 30-ng injections. The 32S/33S and 33S/34S isotope-ratio precision was better than 0.4% RSD for injections of 10 ng (as S) and higher. Detection limits were in the absolute pg range for all measured sulfur isotopes.