Platzner I, Ehrlich S, Halicz L
Geological Survey of Israel, Jerusalem.
Fresenius J Anal Chem. 2001 Jul;370(5):624-8. doi: 10.1007/s002160100875.
The capability of a second-generation Nu Instruments multiple collector inductively coupled plasma mass spectrometer (MC-ICP-MS) has been evaluated for precise and accurate isotope-ratio determinations of lead. Essentially the mass spectrometer is a double-focusing instrument of Nier-Johnson analyzer geometry equipped with a newly designed variable-dispersion ion optical device, enabling the measured ion beams to be focused into a fixed array of Faraday collectors and an ion-counting assembly. NIST SRM Pb 981, 982, and 983 isotopic standards were used. Addition of thallium to the lead standards and subsequent simultaneous measurement of the thallium and lead isotopes enabled correction for mass discrimination, by use of the exponential correction law and 205Tl/203Tl = 2.3875. Six measurements of SRM Pb-982 furnished the results 206Pb/204Pb = 36.7326(68), 207Pb/204Pb = 17.1543(30), 208Pb/204Pb = 36.7249(69), 207Pb/206Pb = 0.46700(1), and 208Pb/206Pb = 0.99979(2); the NIST-certified values were 36.738(37), 17.159(25), 36.744(50), 0.46707(20), and 1.00016(36), respectively. Direct isotope lead analysis in silicates can be performed without any chemical separation. NIST SRM 610 glass was dissolved and introduced into the MC-ICP-MS by means of a micro concentric nebulizer. The ratios observed were in excellent agreement with previously reported data obtained by TIMS and laser ablation MC-ICP-MS, despite the high Ca/Pb concentration ratio (200/1) and the presence of many other elements at levels comparable with that of lead. Approximately 0.2 microg lead are sufficient for isotope analysis with ratio uncertainties between 240 and 530 ppm.
对第二代Nu Instruments多接收电感耦合等离子体质谱仪(MC-ICP-MS)测定铅同位素比的精密度和准确度进行了评估。该质谱仪本质上是一台尼尔-约翰逊分析器几何结构的双聚焦仪器,配备了新设计的可变色散离子光学装置,能将测量的离子束聚焦到固定排列的法拉第杯和一个离子计数组件中。使用了NIST SRM Pb 981、982和983同位素标准物质。向铅标准物质中加入铊,随后同时测量铊和铅同位素,通过使用指数校正定律以及205Tl/203Tl = 2.3875来校正质量歧视。对SRM Pb-982进行的六次测量给出了以下结果:206Pb/204Pb = 36.7326(68)、207Pb/204Pb = 17.1543(30)、208Pb/204Pb = 36.7249(69)、207Pb/206Pb = 0.46700(1)以及208Pb/206Pb = 0.99979(2);而NIST认证值分别为36.7