Ruegsegger D R, Lerude S D, Lyle D
Radiology. 1979 Nov;133(2):483-9. doi: 10.1148/133.2.483.
Dose distributions in electron-beam arc therapy were investigated using the electron beam from a Brown Boveri 45 MeV betatron. This variable isocenter machine is equipped with an angular speed control device that synchronizes the radiation dose rate with the rotation speed about the isocenter. An Alderson Rando phantom was used with Kodak RP/V film to measure isodensity distributions that were then converted to isodose curves. The effects of varying field size, isocenter depth, and air gap between collimator and phantom were examined, along with changes in the dose distributions produced by wax bolus in part of the arc.
利用布朗·博韦里45兆电子伏特电子感应加速器产生的电子束,对电子束弧形治疗中的剂量分布进行了研究。这台可变等中心机器配备了一个角速度控制装置,该装置能使辐射剂量率与围绕等中心的旋转速度同步。使用阿尔德森·兰多人体模型和柯达RP/V胶片来测量等密度分布,然后将其转换为等剂量曲线。研究了不同射野大小、等中心深度以及准直器与人体模型之间气隙的影响,同时还研究了弧形部分中蜡块所产生的剂量分布变化。