Abstract

The analytical method for calculation of electron dose distributions described in part I was experimentally verified using a 42-MeV betatron. For this, a comparison was made between the calculated and the measured dose distributions obtained with a homogeneous or inhomogeneous water or polystyrol phantom with plane or with uneven surfaces, the inhomogeneities consisting in cavities and cork intercalations. Thus, the accuracy of the algorithm and of the consideration for inhomogeneities and irregular surfaces is demonstrated, and limitations of the method's applicability are shown. Some examples of calculated dose distributions in coplanar irradiation techniques using solely electron beams or else a combined treatment with photons and electrons are exhibiting the particular bearing of electronic data processing on the treatment planning in depth therapy with electrons in the energy range above 20 MeV.

摘要

第一部分所述的电子剂量分布计算分析方法,使用一台42兆电子伏的电子感应加速器进行了实验验证。为此,对用均匀或非均匀水模或聚苯乙烯模(表面为平面或不平整)获得的计算剂量分布与测量剂量分布进行了比较,非均匀性包括腔体和软木夹层。由此证明了算法以及对非均匀性和不规则表面考虑的准确性,并展示了该方法适用性的局限性。一些仅使用电子束的共面照射技术或光子与电子联合治疗的计算剂量分布示例,展现了电子数据处理对能量在20兆电子伏以上的电子深部治疗计划的特殊意义。