Leng C T, Crosson J E, Calkins H, Berger R D
Department of Medicine, The Johns Hopkins Medical Institutions, Baltimore, Maryland, USA.
Pacing Clin Electrophysiol. 2001 Aug;24(8 Pt 1):1291-2. doi: 10.1046/j.1460-9592.2001.01291.x.
The authors devised a nonthoracotomy defibrillation system for a patient with a prosthetic tricuspid valve using existing technology and previously established implantation techniques. Their lead configuration deviates substantially from existing designs in its primary use of a coronary sinus defibrillation coil and a left-sided subcutaneous array to distribute current across the ventricular myocardium.