To evaluate the effectiveness of electrocardiographically (ECG)-gated retrospective image reconstruction for multi-detector row computed tomographic (CT) coronary angiography in reducing cardiac motion artifacts and to evaluate the influence of heart rate on cardiac image quality.
Sixty-five patients with different heart rates underwent coronary CT angiography. Raw helical CT data and ECG tracings were combined to retrospectively reconstruct at the defined consecutive z position with a temporal resolution of 250 msec per section. The starting points of the reconstruction were chosen between 30% and 80% of the R-R intervals. The relationships between heart rate, trigger delay, and image quality were analyzed.
Optimal image quality was achieved with a 50% trigger delay for the right coronary artery and 60% for the left circumflex coronary artery. Optimal image quality for the left anterior descending coronary artery was equally obtained at 50% and 60% triggering. A significant negative correlation was observed between heart rate and image quality (P <.05). The best image quality was achieved when the heart rate was less than 74.5 beats per minute.
To achieve high image quality, the heart rate should be sufficiently slow. Selection of appropriate trigger delays and a decreasing heart rate are effective to reduce cardiac motion artifacts.
评估心电图(ECG)门控回顾性图像重建在多排螺旋计算机断层扫描(CT)冠状动脉造影中减少心脏运动伪影的有效性,并评估心率对心脏图像质量的影响。
65例心率不同的患者接受了冠状动脉CT血管造影。将原始螺旋CT数据与心电图描记相结合,以250毫秒/层的时间分辨率在定义的连续z位置进行回顾性重建。重建的起始点在R-R间期的30%至80%之间选择。分析心率、触发延迟和图像质量之间的关系。
右冠状动脉触发延迟50%、左旋支冠状动脉触发延迟60%时可获得最佳图像质量。左前降支冠状动脉在触发延迟50%和60%时均可获得最佳图像质量。观察到心率与图像质量之间存在显著负相关(P<.05)。心率低于每分钟74.5次时可获得最佳图像质量。
为获得高图像质量,心率应足够慢。选择合适的触发延迟和降低心率可有效减少心脏运动伪影。