Stoll M H, Bakker K, Nobbe G H, Haese R R
Netherlands Institute for Sea Research, Den Burg.
Anal Chem. 2001 Sep 1;73(17):4111-6. doi: 10.1021/ac010303r.
A rapid, continuous-flow determination of total inorganic carbon (TIC) in seawater samples is presented. The method runs on an autoanalyzer Traacs 800 spectrophotometric system and is calibrated versus certified reference materials readily available. A typical analysis speed of 45 samples/h can be reached with an accuracy of 2-3 microM and a precision of approximately 2.5 microM. The analysis requires only a small amount of sample and is thus ideally suited for pore water samples and samples taken from cultures where sample volume is at a premium. The speed of the analysis makes mapping of oceanic surface water characteristics possible. Potential interference of sulfide in anoxic (e.g., pore water) samples can be masked by the addition of a hydrogen peroxide step. Although the latter is a strong oxidative reagent, no significant effect on TIC concentration due to oxidation of (labile) organic matter could be found.
本文介绍了一种快速、连续流动测定海水样品中总无机碳(TIC)的方法。该方法在自动分析仪Traacs 800分光光度系统上运行,并与现成的有证标准物质进行校准。典型的分析速度可达每小时45个样品,准确度为2 - 3微摩尔,精密度约为2.5微摩尔。该分析仅需少量样品,因此非常适合孔隙水样品以及取自培养物的样品,因为这些样品的体积非常宝贵。分析速度使得绘制海洋表层水特征图成为可能。通过添加过氧化氢步骤,可以掩盖缺氧(如孔隙水)样品中硫化物的潜在干扰。尽管过氧化氢是一种强氧化试剂,但未发现由于(不稳定)有机物氧化而对TIC浓度产生显著影响。