Fluorometric determination of ethanol in liquor samples by flow-injection analysis using an immobilized enzyme-reactor column with packing prepared by coupling alcohol oxidase and peroxidase onto chitosan beads.

作者信息

Taniai T, Sukuragawa A, Okutani T

机构信息

Nihon University, College of Science and Technology, Department of Materials and Applied Chemistry, Chiyoda, Tokyo, Japan.

出版信息

J AOAC Int. 2001 Sep-Oct;84(5):1475-83.

Abstract

A flow-injection system was developed for the determination of ethanol with an immobilized enzyme-reactor column. This system, which consisted of hand-made reactor columns packed with alcohol oxidase and horseradish peroxidase immobilized onto chitosan beads, and a fluorometric detector, was applied to the determination of ethanol in liquor samples. Under the recommended conditions, the ethanol, which was present in the pretreated samples, was converted to hydrogen peroxide when it was passed through the immobilized alcohol oxidase (AOD) column with 0.1 mol/dm3 phosphate buffer (pH 7.0). A sample can be analyzed with this system in <10 min. The calibration curve for ethanol was linear from 2.0 to 0.1 mg/dm3. The determination limit, which was defined by the difference between the sample peak and blank peak, was estimated to be 50 microg/dm3 for ethanol. Interferences from some substances present in actual liquor samples decreased the analytical response and activity of the immobilized AOD-reactor column, but they were removed by dilution and pretreatment with an octyldecylsilane cartridge.