Furman S, Escher D J
Ann Thorac Surg. 1975 Sep;20(3):326-38. doi: 10.1016/s0003-4975(10)64225-7.
Six hundred nineteen patients have been followed by remote monitoring of pacemaker function using ECG and rate or rate alone; 278 of 280 have had battery exhaustion or electronic failure demonstrated. Ten percent of exhausted pacemakers failed prior to the average longevity of the particular model, and 32% (89 of 280) exceeded 36 months' longevity; of these, 13% (37 of 280) lasted more than 40 months and 4.6% (13 of 280) exceeded 50 months. The error rate is 0.7% (2 of 280). With pulse generator longevity increasing, monitoring is done less frequently during the first 2 years, then calls are made weekly after 24 months.
已通过使用心电图和心率或仅通过心率对619名患者进行起搏器功能的远程监测;280例中有278例已证实存在电池耗尽或电子故障。10%的耗尽型起搏器在特定型号的平均使用寿命之前出现故障,32%(280例中的89例)超过36个月的使用寿命;其中,13%(280例中的37例)持续超过40个月,4.6%(280例中的13例)超过50个月。错误率为0.7%(280例中的2例)。随着脉冲发生器使用寿命的延长,在最初2年内监测频率较低,24个月后每周进行一次呼叫。