Blake D A, Jones R M, Blake R C, Pavlov A R, Darwish I A, Yu H
Department of Ophthalmology, Tulane University Health Sciences Center, New Orleans, LA 70112, USA.
Biosens Bioelectron. 2001 Dec;16(9-12):799-809. doi: 10.1016/s0956-5663(01)00223-8.
Competitive immunoassays for Cd(II), Co(II), Pb(II) and U(VI) were developed using identical reagents in two different assay formats, a competitive microwell format and an immunosensor format with the KinExA 3000. Four different monoclonal antibodies specific for complexes of EDTA-Cd(II), DTPA-Co(II), 2,9-dicarboxyl-1,10-phenanthroline-U(VI), or cyclohexyl-DTPA-Pb(II) were incubated with the appropriate soluble metal-chelate complex. In the microwell assay format, the immobilized version of the metal-chelate complex was present simultaneously in the assay mixture. In the KinExA format, the antibody was allowed to pre-equilibrate with the soluble metal-chelate complex, then the incubation mixture was rapidly passed through a microcolumn containing the immobilized metal-chelate complex. In all four assays, the KinExA format yielded an assay with 10-1000-fold greater sensitivity. The enhanced sensitivity of the KinExA format is most likely due to the differences in the affinity of the monoclonal antibodies for the soluble versus the immobilized metal-chelate complex. The KinExA 3000 instrument and the Cd(II)-specific antibody were used to construct a prototype assay that could correctly assess the concentration of cadmium spiked into a groundwater sample. Mean analytical recovery of added Cd(II) was 114.25+/-11.37%. The precision of the assay was satisfactory; coefficients of variation were 0.81-7.77% and 3.62-14.16% for within run and between run precision, respectively.
使用相同试剂以两种不同的检测形式开发了用于检测镉(II)、钴(II)、铅(II)和铀(VI)的竞争性免疫分析方法,一种是竞争性微孔板形式,另一种是使用KinExA 3000的免疫传感器形式。将四种分别对乙二胺四乙酸 - 镉(II)、二乙三胺五乙酸 - 钴(II)、2,9 - 二羧基 - 1,10 - 菲咯啉 - 铀(VI)或环己基二乙三胺五乙酸 - 铅(II)复合物具有特异性的不同单克隆抗体与相应的可溶性金属螯合物复合物一起孵育。在微孔板检测形式中,金属螯合物复合物的固定化形式同时存在于检测混合物中。在KinExA形式中,使抗体与可溶性金属螯合物复合物预平衡,然后将孵育混合物快速通过含有固定化金属螯合物复合物的微柱。在所有四种检测中,KinExA形式产生的检测灵敏度提高了10 - 1000倍。KinExA形式灵敏度增强最可能是由于单克隆抗体对可溶性与固定化金属螯合物复合物的亲和力存在差异。使用KinExA 3000仪器和镉(II)特异性抗体构建了一种原型检测方法,该方法能够正确评估添加到地下水样品中的镉浓度。添加镉(II)的平均分析回收率为114.25±11.37%。该检测的精密度令人满意;批内精密度的变异系数为0.81 - 7.77%,批间精密度的变异系数分别为3.62 - 14.16%。