Hesse Y, Freissler K A, Lang G K
Universitäts-Augenklinik, Prittwitzstrasse 43, 89075 Ulm.
Klin Monbl Augenheilkd. 2001 Oct;218(10):682-7. doi: 10.1055/s-2001-18391.
Evaluation of the surface properties of two models of one-piece foldable acrylic intraocular lenses (IOLs) before and after folding and passing the IOLs through an injector.
We studied two foldable IOL models made of a polymethacrylate/2-hydroxyethylmethacrylate copolymer: EasAcryl 1(R) (Chiron) and Raysoft(R) 547H (Rayner). Twelve IOLs of each design with different diopters were examined. Light and scanning electron microscopy were performed before and after IOL folding and passing the IOLs through a Micra G103 injector (Chiron Adatomed) and an IC-2BU inserter cartridge (Chiron Microsert II).
All IOLs presented a homogeneous surface and smoothly finished edges of haptics and optics. The inner edges of all IOL haptics and the optic edges of some IOLs had ridges at high magnifications. No IOL demonstrated any folding defects of the optic such as a folding line or scratches. Three of 16 folded IOLs showed defects or cracking of the IOL haptics due to incomplete placement in the inserter cartridge.
The two models of soft acrylic intraocular lenses showed excellent surface quality. After folding and injection process no alteration of the optics was detected in any of the IOLs. However, the IOLs need careful placement in the inserter cartridge to avoid haptic damage.
评估两种一体式可折叠丙烯酸人工晶状体(IOL)模型在折叠前、折叠后以及通过注射器后的表面特性。
我们研究了由聚甲基丙烯酸酯/甲基丙烯酸2-羟乙酯共聚物制成的两种可折叠IOL模型:EasAcryl 1(Chiron公司)和Raysoft 547H(Rayner公司)。对每种设计的12个不同屈光度的IOL进行了检查。在IOL折叠前、折叠后以及通过Micra G103注射器(Chiron Adatomed公司)和IC-2BU插入器盒(Chiron Microsert II)后,进行了光学显微镜和扫描电子显微镜检查。
所有IOL均呈现出均匀的表面,其襻和光学部边缘加工光滑。在高倍放大下,所有IOL襻的内边缘以及部分IOL的光学部边缘有嵴。没有IOL出现光学部的任何折叠缺陷,如折叠线或划痕。16个折叠后的IOL中有3个由于未完全放置在插入器盒中而出现襻的缺陷或破裂。
两种软性丙烯酸人工晶状体模型均显示出优异的表面质量。在折叠和注射过程后,未在任何IOL中检测到光学部的改变。然而,IOL在插入器盒中需要小心放置,以避免襻受损。