Omar O, Mamalis N, Veiga J, Tamura M, Olson R J
John A. Moran Eye Center, University of Utah, Salt Lake City 84132, USA.
Ophthalmology. 1996 Jul;103(7):1124-9. doi: 10.1016/s0161-6420(96)30557-5.
To evaluate the surface characteristics of commonly used, small-incision, intraocular lenses (IOLs).
Representative samples of five groups of foldable IOLs (4 silicone and 1 acrylic) underwent surface and edge-finish examination using a slit lamp. The IOLs were folded using a folding block and forceps. All the IOLs then were examined using a scanning electron microscope. A one-piece polymethylmethacrylate IOL was used for comparing surface-finish characteristics. The IOLs were examined for optic surface quality, edge finish, haptic, haptic/optic junction, and possible post-folding modifications.
Slit-lamp evaluation of the surface quality of all of the silicone lenses demonstrated a smooth finish of the optic surface, edge, and haptics. Scanning electron microscopic analysis of the IOLs demonstrated adequately finished haptics or footplates and optics. Excess molding flash was seen on the edges of the some of the silicone IOLs, and no molding flash was observed on others. The acrylic IOL had a somewhat sharper optic edge. Irregular finish of the haptic/optic junctions of some of the IOLs (both silicone and acrylic) was noted.
Currently available foldable IOLs have demonstrated adequate lens finish. However, irregularities of the haptic/optic junctions and molding flash are present on most IOLs evaluated, indicating room for improvement in the finish of foldable IOLs. Phacoemulsification with capsular bag IOL placement may decrease the clinical significance of these relatively subtle lens finish irregularities.
评估常用小切口人工晶状体(IOL)的表面特性。
使用裂隙灯对五组可折叠IOL(4种硅酮材质和1种丙烯酸酯材质)的代表性样本进行表面和边缘处理检查。使用折叠块和镊子对IOL进行折叠。然后使用扫描电子显微镜对所有IOL进行检查。使用一片式聚甲基丙烯酸甲酯IOL来比较表面处理特性。对IOL的光学表面质量、边缘处理、襻、襻/光学部连接处以及折叠后可能的改变进行检查。
裂隙灯对所有硅酮材质IOL的表面质量评估显示,光学表面、边缘和襻表面光滑。IOL的扫描电子显微镜分析显示襻或脚板以及光学部处理良好。在一些硅酮IOL的边缘可见多余的成型飞边,而其他IOL未观察到成型飞边。丙烯酸酯IOL的光学边缘稍尖锐。注意到一些IOL(硅酮和丙烯酸酯材质)的襻/光学部连接处表面不平整。
目前可用的可折叠IOL已显示出足够的镜片表面处理。然而,在大多数评估的IOL上存在襻/光学部连接处不平整和成型飞边,表明可折叠IOL的表面处理仍有改进空间。囊袋内植入IOL的超声乳化术可能会降低这些相对细微的镜片表面处理不平整的临床意义。