Suppr超能文献

新生儿重症监护病房中插管患者沙丁胺醇给药实践的调查。

A survey of albuterol administration practices in intubated patients in the neonatal intensive care unit.

作者信息

Ballard Julie, Lugo Ralph A, Salyer John W

机构信息

Respiratory Care Department, Primary Children's Medical Center, 100 North Medical Drive, Salt Lake City UT 84113-1100, USA.

出版信息

Respir Care. 2002 Jan;47(1):31-8.

Abstract

INTRODUCTION

Aerosolized albuterol is commonly used in the treatment of neonatal respiratory illnesses. Clinical and in vitro studies have identified numerous factors that affect aerosol drug delivery during neonatal mechanical ventilation, including the choice of metered-dose inhaler (MDI) or nebulizer, the use of a holding chamber, time between actuations, the volume of nebulized solution, and the position and placement of the nebulizer or MDI. Because there is no consensus on the optimal method of administration, there is probably substantial variability among institutions in how aerosolized albuterol is administered to mechanically ventilated infants in the neonatal intensive care unit (NICU).

OBJECTIVE

Survey academic medical centers in the United States regarding their practices of administering aerosolized albuterol to intubated newborns in the NICU.

METHODS

A survey instrument was developed that queried 18 aspects of albuterol administration in mechanically ventilated infants, including the frequency of MDI and nebulizer use, the average and maximum dose, the time between MDI actuations and following the final actuation, the use of a holding chamber, and the placement location of the holding chamber or nebulizer. Respiratory therapists and respiratory therapy managers having direct knowledge of neonatal clinical practices in their neonatal fellowship program NICUs were surveyed via telephone. Those who did not respond via telephone were surveyed via fax.

RESULTS

Eighty institutions were surveyed and there were 68 respondents (85% response rate). Responders averaged 35 +/- 13 NICU beds and 11 +/- 5 ventilators/d. Nineteen percent of the respondents reported administering albuterol via MDI 100% of the time; 22% use MDIs 75-99% of the time; 9% use MDIs 50-74% of the time; 4% use MDIs 25-49% of the time; and 43% never use MDIs to deliver albuterol. The average dose via MDI was: 1 puff: 30%; 2 puffs: 65%; and 4 puffs: 5%. The maximum dose via MDI was: 2 puffs: 30%; 3 puffs: 14%; 4 puffs: 36%; 6 puffs: 11%; and 8 puffs: 6%. Thirty-one percent of the respondents place the holding chamber in-line with the ventilator circuit, 56% administer the aerosol via manual ventilation, and 13% use both methods. Fifty-six percent place the in-line holding chamber between the endotracheal tube and ventilator circuit, and the other 44% place the in-line holding chamber in the inspiratory limb. The time between MDI actuations depended on whether the holding chamber was placed in-line or the aerosol was administered via manual ventilation (MV): < or = 0.5 min: 18% in-line and 28% MV; 1 min: 47% in-line and 43% MV; 2 min: 6% in-line and 4% MV; 3 min: 6% in-line and 0% MV. Eighty-three percent of respondents indicated that dead space introduced by a holding chamber/spacer was not a concern. Forty-three percent use nebulizers exclusively to administer albuterol to mechanically ventilated patients. Seventy-four percent of centers that nebulize albuterol use a dose of 1.25-2.5 mg. Eighty-eight percent of the surveyed institutions place nebulizers in-line with the ventilator circuit, and the other 12% use manual ventilation to administer the nebulized aerosol. Of those that use in-line nebulization, 95% place the nebulizer in the inspiratory limb of the circuit, and the other 5% place the nebulizer between the endotracheal tube and circuit Y-piece. Among centers that place the nebulizer in the inspiratory limb, 52% place it adjacent to the circuit Y-piece, 36% place it midway upstream in the inspiratory limb, and 12% place it near the humidifier.

CONCLUSION

There is substantial variability among NICUs in albuterol administration to mechanically ventilated infants, with the majority of institutions now administering albuterol via MDI.

摘要

引言

雾化沙丁胺醇常用于治疗新生儿呼吸系统疾病。临床和体外研究已经确定了许多影响新生儿机械通气期间雾化药物递送的因素,包括定量吸入器(MDI)或雾化器的选择、储雾罐的使用、按压间隔时间、雾化溶液的体积以及雾化器或MDI的位置和放置方式。由于对于最佳给药方法尚无共识,新生儿重症监护病房(NICU)中各机构在对机械通气婴儿使用雾化沙丁胺醇的方式上可能存在很大差异。

目的

调查美国学术性医疗中心对NICU中插管新生儿使用雾化沙丁胺醇的做法。

方法

开发了一份调查问卷,询问了机械通气婴儿使用沙丁胺醇的18个方面,包括MDI和雾化器的使用频率、平均和最大剂量、MDI按压间隔时间以及最后一次按压之后的时间、储雾罐的使用以及储雾罐或雾化器的放置位置。通过电话对在其新生儿专科培训项目NICU中直接了解新生儿临床实践的呼吸治疗师和呼吸治疗管理人员进行了调查。未通过电话回复的人员通过传真进行了调查。

结果

共调查了80家机构,有68名受访者(回复率85%)。受访者的NICU平均床位为35±13张,呼吸机平均每天使用11±5台。19%的受访者报告100%的时间通过MDI使用沙丁胺醇;22%在75 - 99%的时间使用MDI;9%在50 - 74%的时间使用MDI;4%在25 - 49%的时间使用MDI;43%从未使用MDI来递送沙丁胺醇。通过MDI的平均剂量为:1喷:30%;2喷:65%;4喷:5%。通过MDI的最大剂量为:2喷:30%;3喷:14%;4喷:36%;6喷:11%;8喷:6%。31%的受访者将储雾罐与呼吸机回路串联放置,56%通过手动通气给予气雾剂,13%同时使用这两种方法。56%将串联储雾罐放置在气管内导管和呼吸机回路之间,另外44%将串联储雾罐放置在吸气支路上。MDI按压间隔时间取决于储雾罐是否串联放置或气雾剂是否通过手动通气(MV)给予:≤0.5分钟:串联放置时为18%,手动通气时为28%;1分钟:串联放置时为47%,手动通气时为43%;2分钟:串联放置时为6%,手动通气时为4%;3分钟:串联放置时为6%,手动通气时为0%。83%的受访者表示储雾罐/储物器引入的死腔不是问题。43%仅使用雾化器对机械通气患者给予沙丁胺醇。74%雾化沙丁胺醇的中心使用剂量为1.25 - 2.5毫克。88%的被调查机构将雾化器与呼吸机回路串联放置,另外12%使用手动通气给予雾化气雾剂。在使用串联雾化的机构中,95%将雾化器放置在回路的吸气支路上,另外5%将雾化器放置在气管内导管和回路Y形接头之间。在将雾化器放置在吸气支路的中心中,52%将其放置在靠近回路Y形接头处,36%将其放置在吸气支路的上游中间位置,12%将其放置在靠近加湿器处。

结论

NICU在对机械通气婴儿使用沙丁胺醇的方式上存在很大差异,现在大多数机构通过MDI使用沙丁胺醇。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验