Diekmann F, Diekmann S, Bick U, Rogalla P, Blohmer J U, Winzer K J, Hamm B
Institut für Radiologie, Universitätsklinikum Charité, Berlin, Germany.
Rofo. 2002 Mar;174(3):297-300. doi: 10.1055/s-2002-20603.
To evaluate the conspicuity of microcalcifications in magnified mammographic views of preparations obtained with full field digital mammography (FFDM), film-screen mammography (FSM), and the DIMA technique.
Twelve preparations were examined by FFDM and FSM using 1.8 x magnification and DIMA using 7 x magnification. Parameter settings were identical for all three techniques. The number of visible microcalcifications was then determined for each modality by three radiologists. As far as possible, all preparations were X-rayed at 22 kV and 10 mAS.
Altogether 9705 calcifications were counted (DIMA: 1609/1542/1534; FFDM: 1020/753/881; FSM: 901/643/822). The total number of microcalcifications identified with the DIMA technique was 4685 as compared to 2654 with FFDM and 2366 with FSM. The calcifications counted with FFDM and FSM thus corresponded to 56.6 % and 50.5 %, respectively, of those identified with DIMA. The differences between the groups were statistically significant (F-Test, p < 0.05).
Significantly more calcifications are identified when magnified mammographic views of preparations containing microcalcifications are obtained with the DIMA technique compared to FFDM or FSM. FFDM depicts markedly more calcifications than FSM. This means one should increase spatial resolution. Digital mammography offers the potential for improved visualization of microcalcifications with advanced applications.
评估在使用全视野数字乳腺摄影(FFDM)、屏-片乳腺摄影(FSM)和数字图像微钙化分析(DIMA)技术获得的放大乳腺摄影图像中微钙化的显示清晰度。
对12份标本分别采用FFDM和FSM进行1.8倍放大检查,采用DIMA进行7倍放大检查。三种技术的参数设置相同。然后由三位放射科医生确定每种检查方式下可见微钙化的数量。所有标本尽可能在22 kV和10 mAS条件下进行X线摄影。
总共计数了9705处钙化(DIMA:1609/1542/1534;FFDM:1020/753/881;FSM:901/643/822)。与FFDM的2654处和FSM的2366处相比,DIMA技术识别出的微钙化总数为4685处。因此,FFDM和FSM计数的钙化分别相当于DIMA识别出的钙化的56.6%和50.5%。各组之间的差异具有统计学意义(F检验,p < 0.05)。
与FFDM或FSM相比,使用DIMA技术获得含有微钙化标本的放大乳腺摄影图像时,识别出的钙化明显更多。FFDM显示的钙化明显多于FSM。这意味着应提高空间分辨率。数字乳腺摄影通过先进的应用为改善微钙化的可视化提供了潜力。