Rock C, Kotsianos D, Linsenmaier U, Fischer T, Brandl R, Vill F, Wirth S, Kaltschmidt R, Euler E, Pfeifer K J, Reiser M
Institut für Klinische Radiologie, Klinikum der Universität München, Innenstadt, Germany.
Rofo. 2002 Feb;174(2):170-6. doi: 10.1055/s-2002-20107.
Evaluation of 3D-CT imaging of the axial skeleton and different joints of the lower and upper extremities with a new dedicated CT system (ISO-C-3D) based on a mobile isocentric C-arm image amplifier.
27 cadaveric specimes of different joints of the lower and upper extremities and of the spinal column were examined with 3D-CT imaging (ISO-C-3D). All images were evaluated by 3 radiologists for image quality using a semiquantitative score (score value 1: poor quality; score value 4: excellent quality). In addition, dose measurements and measurements of high contrast resolution were performed in comparison to conventional and low-dose spiral CT using a high contrast phantom (Catphan, Phantom Laboratories).
Adequate image quality (mean score values 3 - 4) could be achieved with an applied dose comparable to low-dose CT in smaller joints such as wrist, elbow, ankle and knee. A remarkably inferior image quality resulted in imaging of the hip, lumbar and thoracic spine (mean score values 2 - 3) in spite of almost doubling the dose (dose increased by 85 percent). The image quality of shoulder examinations was insufficient (mean score value 1). Phantom studies showed a high-contrast resolution comparable to helical CT in the xy-axis (9 lp/cm).
Preliminary results show, that image quality of C-arm-based CT-imaging (ISO-C-3D) seems to be adequate in smaller joints. ISO-C-3D images of the hip and axial skeleton show a decreased image quality, which does not seem to be sufficient for diagnosing subtle fractures.
使用基于移动等中心C型臂影像增强器的新型专用CT系统(ISO-C-3D)对轴骨以及上下肢不同关节进行三维CT成像评估。
对27具上下肢不同关节及脊柱的尸体标本进行三维CT成像(ISO-C-3D)检查。3名放射科医生使用半定量评分(评分值1:质量差;评分值4:质量优)对所有图像的质量进行评估。此外,使用高对比度体模(Catphan,体模实验室)与传统CT和低剂量螺旋CT相比,进行剂量测量和高对比度分辨率测量。
在腕关节、肘关节、踝关节和膝关节等较小关节中,使用与低剂量CT相当的剂量即可获得足够的图像质量(平均评分值3 - 4)。尽管剂量几乎翻倍(增加了85%),但在髋关节、腰椎和胸椎成像中图像质量明显较差(平均评分值2 - 3)。肩关节检查的图像质量不足(平均评分值1)。体模研究表明,在xy轴上高对比度分辨率与螺旋CT相当(9 lp/cm)。
初步结果表明,基于C型臂的CT成像(ISO-C-3D)在较小关节中的图像质量似乎足够。髋关节和轴骨的ISO-C-3D图像显示图像质量下降,这似乎不足以诊断细微骨折。