Suppr超能文献

锥形束 O 臂成像系统的剂量学特性描述。

Dosimetric characterization of a cone-beam O-arm imaging system.

机构信息

Medical Physics Section, Department of Radiology, University of Minnesota, Minneapolis, MN 55455, USA.

出版信息

J Xray Sci Technol. 2009;17(4):305-17. doi: 10.3233/XST-2009-0231.

Abstract

This study compared patient dose and image quality of a mobile O-arm cone beam imaging system in the 3D scan acquisition mode to those of a 64 slice Computed Tomography (CT) imaging system. The investigation included patient dose, scattered radiation, and image quality measurements. The patient dose was measured using a 0.6 cc Farmer ion chamber and 30 cm long Computed Tomography (CT) head and body polymethylmethacrylate (PMMA) phantoms. The results show that under identical radiographic techniques (kVp, mAs, etc.) and with the same scan length, the O-arm in 3D scan acquisition mode delivers approximately half the radiation dose of a 64 slice CT scanner. Scattered radiation was measured at several locations around the O-arm, at 1 m, 2 m and 3 m distances in 3D CT scan acquisition mode with a RadCal 10 x 5-180 pancake ion chamber using a 30 cm long CT body phantom as the source of scatter. Similar measurements were made in a 64 slice CT scanner. The data demonstrate that scattered radiation from the O-arm to personnel involved in a clinical procedure is comparable to that from a 64 slice CT scanner. Image quality was compared by exposing a CATPHAN phantom to comparable doses in both the O-arm and the CT scanner. The resultant images were then evaluated for modulation transfer function (MTF), high-contrast spatial resolution, and low contrast sensitivity for clinical application purpose. The O-arm shows comparable high contrast to the CT (7 lp/cm vs. 8 lp/cm). The low contrast in the O-arm is not visible due to fixed pattern noise. For image guided surgery applications where the location of a structure is emphasized over a survey of all image details, the O-arm has some advantages due to wide radiation beam coverage and lower patient dose. The image quality of the O-arm needs significant improvement for other clinical applications where high image quality is desired.

摘要

本研究比较了移动 O 臂锥形束成像系统在 3D 扫描采集模式下与 64 层 CT 成像系统的患者剂量和图像质量。该研究包括患者剂量、散射辐射和图像质量测量。患者剂量使用 0.6cc 的 Farmer 离子室和 30cm 长的 CT 头部和体部聚甲基丙烯酸甲酯(PMMA)体模进行测量。结果表明,在相同的射线照相技术(kVp、mAs 等)和相同的扫描长度下,O 臂在 3D 扫描采集模式下提供的辐射剂量约为 64 层 CT 扫描仪的一半。在 3D CT 扫描采集模式下,使用 RadCal 10 x 5-180 平面电离室在 O 臂周围的几个位置、1m、2m 和 3m 距离处测量散射辐射,使用 30cm 长的 CT 体模作为散射源。在 64 层 CT 扫描仪中进行了类似的测量。数据表明,在临床操作过程中,O 臂对人员的散射辐射与 64 层 CT 扫描仪相当。通过在 O 臂和 CT 扫描仪中以可比剂量使 CATPHAN 体模曝光,比较了图像质量。然后,根据调制传递函数(MTF)、高对比度空间分辨率和低对比度灵敏度对图像进行评估,以用于临床应用。O 臂的高对比度与 CT 相当(7lp/cm 对 8lp/cm)。由于固定模式噪声,O 臂中的低对比度不可见。对于图像引导手术应用,由于结构位置强调而不是所有图像细节的调查,O 臂具有一些优势,因为宽辐射束覆盖范围和低患者剂量。对于需要高图像质量的其他临床应用,O 臂的图像质量需要显著提高。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验