使用光学反射换能器对体外循环中的氧饱和度进行实时监测。

Real time monitoring of oxygen saturation in extracorporeal circulation using an optical reflectance transducer.

作者信息

Gamero F U, Ushizima M R, Cestari I A

机构信息

Department of Physiology and Biophysics, ICB, University of São Paulo, Brazil.

出版信息

Artif Organs. 2001 Nov;25(11):890-4. doi: 10.1046/j.1525-1594.2001.06899.x.

Abstract

A novel optical reflectance transducer for continuous and real time monitoring of oxygen saturation (So2) in extracorporeal tubings during cardiopulmonary bypass (CPB) is described. The optical transducer integrates 2 light-emitting diodes and a photodiode. The transducer is attached to the extracorporeal tubing by a transparent flow through a 3/8 x 3/8 inch connector/cuvette. Four transducers were built and evaluated in vitro in a mock extracorporeal system at a controlled So2. The correlation index (r2) between the So2 measured by the transducers and values determined by a commercial blood gas analyzer was 0.998 with an absolute difference < 0.5%. The performance of the transducer was also evaluated during 16 CPB surgeries in patients with an r2 of 0.873 and an absolute difference < 3.5%. The results obtained demonstrate the applicability of the optical reflectance transducer for monitoring So2 in the tubings of the CPB circuit.

摘要

描述了一种新型光学反射换能器,用于在体外循环(CPB)期间连续实时监测体外管道中的氧饱和度(So2)。该光学换能器集成了2个发光二极管和1个光电二极管。换能器通过一个透明流路,经3/8×3/8英寸连接器/比色皿连接到体外管道。制作了4个换能器,并在体外模拟系统中于受控的So2条件下进行了体外评估。换能器测得的So2与商用血气分析仪测定值之间的相关指数(r2)为0.998,绝对差值<0.5%。在16例CPB手术患者中对换能器的性能也进行了评估,r2为0.873,绝对差值<3.5%。所获得的结果证明了光学反射换能器在监测CPB回路管道中So2方面的适用性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验