影响光电和电磁跟踪系统术中精度及稳定性的因素的定量分析

Quantitative analysis of factors affecting intraoperative precision and stability of optoelectronic and electromagnetic tracking systems.

作者信息

Wagner A, Schicho K, Birkfellner W, Figl M, Seemann R, König F, Kainberger Franz, Ewers R

机构信息

University Hospital of Cranio-Maxillofacial and Oral Surgery, Medical School, University of Vienna, Austria.

出版信息

Med Phys. 2002 May;29(5):905-12. doi: 10.1118/1.1469625.

Abstract

This study aims to provide a quantitative analysis of the factors affecting the actual precision and stability of optoelectronic and electromagnetic tracking systems in computer-aided surgery under real clinical/intraoperative conditions. A "phantom-skull" with five precisely determined reference distances between marker spheres is used for all measurements. Three optoelectronic and one electromagnetic tracking systems are included in this study. The experimental design is divided into three parts: (1) evaluation of serial- and multislice-CT (computed tomography) images of the phantom-skull for the precision of distance measurements by means of navigation software without a digitizer, (2) digitizer measurements under realistic intraoperative conditions with the factors OR-lamp (radiating into the field of view of the digitizer) or/and "handling with ferromagnetic surgical instruments" (in the field of view of the digitizer) and (3) "point-measurements" to analyze the influence of changes in the angle of inclination of the stylus axis. Deviations between reference distances and measured values are statistically investigated by means of analysis of variance. Computerized measurements of distances based on serial-CT data were more precise than based on multislice-CT data. All tracking systems included in this study proved to be considerably less precise under realistic OR conditions when compared to the technical specifications in the manuals of the systems. Changes in the angle of inclination of the stylus axis resulted in deviations of up to 3.40 mm (mean deviations for all systems ranging from 0.49 to 1.42 mm, variances ranging from 0.09 to 1.44 mm), indicating a strong need for improvements of stylus design. The electromagnetic tracking system investigated in this study was not significantly affected by small ferromagnetic surgical instruments.

摘要

本研究旨在对实际临床/手术条件下计算机辅助手术中影响光电和电磁跟踪系统实际精度和稳定性的因素进行定量分析。使用一个在标记球之间有五个精确确定的参考距离的“模拟颅骨”进行所有测量。本研究包括三个光电跟踪系统和一个电磁跟踪系统。实验设计分为三个部分:(1) 通过无数字化仪的导航软件评估模拟颅骨的序列和多层CT(计算机断层扫描)图像用于距离测量的精度;(2) 在实际手术条件下进行数字化仪测量,考虑手术室灯(辐射到数字化仪视野内)或/和“使用铁磁性手术器械操作”(在数字化仪视野内)等因素;(3) 进行“点测量”以分析触笔轴倾斜角度变化的影响。通过方差分析对参考距离和测量值之间的偏差进行统计学研究。基于序列CT数据的距离计算机化测量比基于多层CT数据的测量更精确。与系统手册中的技术规格相比,本研究中包含的所有跟踪系统在实际手术室条件下的精度都明显较低。触笔轴倾斜角度的变化导致偏差高达3.40毫米(所有系统的平均偏差范围为0.49至1.42毫米,方差范围为0.09至1.44毫米),这表明迫切需要改进触笔设计。本研究中调查的电磁跟踪系统不受小型铁磁性手术器械的显著影响。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验