Suppr超能文献

一种用于测量人类和非人类物种皮肤温度敏感性的装置。

A device to measure cutaneous temperature sensitivity in humans and subhuman species.

作者信息

Kenshalo D R, Bergen D C

出版信息

J Appl Physiol. 1975 Dec;39(6):1038-40. doi: 10.1152/jappl.1975.39.6.1038.

Abstract

A device is described to maintain restricted areas of skin at any temperature between 5 and 45 degrees C. Changes in temperature of controlled intensity up to 10 degrees C at rates from 0.03 degrees C to 2 degrees C/s can be delivered in either the warm or cool directions. The stimulator, which is in contact with the skin, is sufficiently simple so that a number of them can be constructed, each with a different contact area up to 18.2 cm2. The current control apparatus that operates a Peltier device in the stimulator is a feedback control system that maintains a precisely controlled temperature at the stimulator-skin interface. Safety features make it suitable and safe for use in human psychophysical studies and subhuman behavioral measurements of temperature sensitivity. Electrostatic shielding makes it compatible with the electronic instruments used in electrophysiological studies of the temperature sense.

摘要

描述了一种用于将皮肤受限区域维持在5至45摄氏度之间任意温度的装置。可以在升温或降温方向上以0.03摄氏度/秒至2摄氏度/秒的速率进行强度可控的高达10摄氏度的温度变化。与皮肤接触的刺激器足够简单,因此可以制造多个,每个刺激器具有不同的接触面积,最大可达18.2平方厘米。在刺激器中操作珀耳帖装置的电流控制装置是一个反馈控制系统,可在刺激器 - 皮肤界面维持精确控制的温度。安全特性使其适用于人体心理物理学研究以及对温度敏感性的亚人类行为测量,且使用安全。静电屏蔽使其与温度感觉的电生理研究中使用的电子仪器兼容。