Lee Kyue-Hyung, Oshima Mitsuko, Motomizu Shoji
Department of Chemistry, Faculty of Science, Okayama University, Japan.
Analyst. 2002 Jun;127(6):769-74. doi: 10.1039/b201210n.
A new on-line flow injection (FI) pre-treatment system using a disk-type chelating resin (5 mm diameter, 0.5 mm thickness) was developed for the simultaneous multi-element determination of trace metals in sea-water samples by inductively coupled plasma mass spectrometry (ICP-MS). A chelating resin possessing an iminodiacetate (IDA) functional group was used for the collection of trace elements and the elimination of alkali and alkaline earth metals in highly concentrated salt solution. A 1 ml volume of a sea-water sample (pH 5.5) was applied to the chelating resin disk. Considering the removal efficiency for Ca, 50 mM ammonium acetate buffer solution (pH 5.5) was chosen as a sample carrier. The enriched trace metals were eluted with 0.1 M nitric acid and the eluate flowed into the ICP-MS system. The processing time for one sample was < 6 min (350 s). One of the important observations is the possibility of working with a low recovery, even lower than 50%. For example, several elements such as Mn, Cr, As, Mo, Ba and U, the recovery of which was < 50% in a batch-wise method, showed good linearity and reproducibility. The proposed method was evaluated by analyzing two kinds of sea-water certified reference materials, CASS-4 and NASS-5. Analytical data for eight heavy metals, V, Mn, Co, Ni, Cu, Mo, Cd and U, obtained from the present study agreed well with the certified values.