Hoetink A E, Faes Th J C, Marcus J T, Kerkkamp H J J, Heethaar R M
Department of Clinical Physics and Informatics, VU Medical Center, Vrije Universiteit, Amsterdam, The Netherlands.
IEEE Trans Med Imaging. 2002 Jun;21(6):653-61. doi: 10.1109/TMI.2002.800582.
Electrical impedance (EI) measurements conducted on the thorax contain useful information about the changes in blood volume that occur in the thorax during the heart cycle. The aim of this paper is to present a new (tomographic-like) method to obtain this relevant information with electrical impedance measurements, using a linear electrode array. This method is tested on three subjects and the results are compared with results, obtained from magnetic resonance cine-images showing the cross-sectional surface area changes of the aorta, the vena cava, the carotid arteries, and the heart. This paper shows that the different sources of the thoracic EI waveform may be separated in time and location on the thoracic surface and that aortic volume changes may be estimated accurately.
对胸部进行的电阻抗(EI)测量包含有关心动周期中胸部血容量变化的有用信息。本文的目的是提出一种新的(类似断层扫描的)方法,使用线性电极阵列通过电阻抗测量来获取此相关信息。该方法在三名受试者身上进行了测试,并将结果与通过磁共振电影图像获得的结果进行了比较,磁共振电影图像显示了主动脉、腔静脉、颈动脉和心脏的横截面积变化。本文表明,胸部EI波形的不同来源可以在胸部表面的时间和位置上分离,并且可以准确估计主动脉容积变化。