School of Electrical and Information Engineering, CARLAB, The University of Sydney, NSW, Australia.
Comput Biol Med. 2012 Nov;42(11):1122-32. doi: 10.1016/j.compbiomed.2012.09.004. Epub 2012 Sep 25.
The conductivity distribution around the thorax is altered during the cardiac cycle due to the blood perfusion, heart contraction and lung inflation. Previous studies showed that these bio-impedance changes are appropriate for non-invasive cardiac function imaging using Electrical Impedance Tomography (EIT) techniques. However, the spatial resolution is presently low. One of the main obstacles in cardiac imaging at the heart location is the large impedance variation of the lungs by respiration and muscles on the dorsal and posterior side of the body. In critical care units there is a potential to insert an internal electrode inside the esophagus directly behind the heart in the same plane of the external electrodes. The aim of the present study is to evaluate different current stimulation and measurement patterns with both external and internal electrodes. Analysis is performed with planar arrangement of 16 electrodes for a simulated 3D cylindrical tank and pig thorax model. In our study we evaluated current injection patterns consisting of adjacent, diagonal, trigonometric, and radial to the internal electrode. The performance of these arrangements was assessed using quantitative methods based on distinguishability, sensitivity and GREIT (Graz consensus Reconstruction algorithm for Electrical Impedance Tomography). Our evaluation shows that an internal electrode configuration based on the trigonometric injection patterns has better performance and improves pixel intensity of the small conductivity changes related to heart near 1.7 times in reconstructed images and also shows more stability with different levels of added noise. For the internal electrode, when we combined radial or adjacent injection with trigonometric injection pattern, we found an improvement in amplitude response. However, the combination of diagonal with trigonometric injection pattern deteriorated the shape deformation (correlation coefficient r=0.344) more than combination of radial and trigonometric injection (correlation coefficient r=0.836) for the perturbations in the area close to the center of the cylinder. We also find that trigonometric stimulation pattern performance is degraded in a realistic thorax model with anatomical asymmetry. For that reason we recommend using internal electrodes only for voltage measurements and as a reference electrode during trigonometric stimulation patterns in practical measurements.
由于血液灌注、心脏收缩和肺部充气,胸部周围的电导率分布在心脏周期中发生变化。先前的研究表明,这些生物阻抗变化适用于使用电阻抗断层成像(EIT)技术进行无创心脏功能成像。然而,目前的空间分辨率较低。在心脏位置进行心脏成像的主要障碍之一是呼吸和身体背部和后部的肌肉对肺部的阻抗变化很大。在重症监护病房,有可能将内部电极插入心脏正后方的食道内,与外部电极位于同一平面。本研究的目的是评估外部和内部电极的不同电流刺激和测量模式。使用模拟 3D 圆柱罐和猪胸部模型的 16 电极平面排列进行分析。在我们的研究中,我们评估了由内部电极相邻、对角、三角和径向的电流注入模式。这些排列的性能是使用基于可区分性、灵敏度和 GREIT(电气阻抗断层成像的 Graz共识重建算法)的定量方法来评估的。我们的评估表明,基于三角注入模式的内部电极配置具有更好的性能,可以将与心脏相关的小电导率变化的像素强度提高近 1.7 倍,并且在不同的噪声水平下也具有更高的稳定性。对于内部电极,当我们将径向或相邻注入与三角注入模式结合使用时,我们发现幅度响应有所提高。然而,当将对角与三角注入模式结合使用时,与将径向与三角注入模式结合使用相比,在靠近圆柱中心的区域中,形状变形的恶化更大(相关系数 r=0.344)。我们还发现,在具有解剖学不对称的现实胸部模型中,三角刺激模式的性能会降低。因此,我们建议仅在实际测量中使用内部电极进行电压测量,并在三角刺激模式下用作参考电极。