Rhodes Anthony, Jasani Bharat, Anderson Elizabeth, Dodson Andrew R, Balaton André J
UK NEQAS-ICC, The Department of Histopathology, University College London Medical School, England.
Am J Clin Pathol. 2002 Sep;118(3):408-17. doi: 10.1309/97WN-W6UX-XJWT-02H2.
Variation in assay sensitivity was studied in more than 90 laboratories that assayed 4formalin-fixed, paraffin-processed breast and ovarian carcinoma cell lines with graded levels of HER-2/neu protein overexpression and known levels of HER-2/neu gene amplification, in addition to breast carcinomas fixed and processed in the laboratories. Main methods were the HercepTest (DAKO, Ely, England) and individualized protocols using a polyclonal antibody and the CB11 clone. While the proportion of laboratories achieving appropriate results with the HercepTest was significantly higher than for participants using other assays, laboratories using other assays showed significant improvement in the second assessment run. The level of agreement in evaluations by 26 laboratories using the HercepTest was excellent on cell lines and tumors and was significantly greater than that achieved by the remaining 41 laboratories using other immunohistochemical methods. While laboratories using the DAKO HercepTest had the highest level of reproducibility in assay sensitivity and evaluation, the significant improvement in results by laboratories using other antibodies in the second assessment run suggests that stringent quality control and an ongoing quality assurance program using a standard reference material have the potential to improve the reliability of immunohistochemical assays for HER-2/neu, regardless of the antibody used.
在90多个实验室中研究了检测灵敏度的差异。这些实验室除了检测各自固定和处理的乳腺癌外,还检测了4种福尔马林固定、石蜡包埋的乳腺癌和卵巢癌细胞系,这些细胞系具有分级水平的HER-2/neu蛋白过表达以及已知水平的HER-2/neu基因扩增。主要方法是HercepTest(丹麦戴科公司,伊利,英国)以及使用多克隆抗体和CB11克隆的个体化方案。虽然使用HercepTest获得适当结果的实验室比例显著高于使用其他检测方法的参与者,但使用其他检测方法的实验室在第二次评估中显示出显著改善。26个使用HercepTest的实验室在细胞系和肿瘤评估中的一致性水平非常好,并且显著高于其余41个使用其他免疫组织化学方法的实验室。虽然使用丹麦戴科公司HercepTest的实验室在检测灵敏度和评估方面具有最高的可重复性水平,但使用其他抗体的实验室在第二次评估中结果的显著改善表明,无论使用何种抗体,严格的质量控制和使用标准参考物质的持续质量保证计划都有可能提高HER-2/neu免疫组织化学检测的可靠性。