Dumont K, Segers R, Vandenberghe S, Van Nooten G, Verdonck P
IBiTech, Institute of Biomedical Technology, Ghent University, Gent, Belgium.
Int J Artif Organs. 2002 Aug;25(8):783-90. doi: 10.1177/039139880202500806.
The aim of this study was to perform combined hydrodynamic and Doppler echocardiographic tests of the Omnicarbon 21 mm cardiac valve in aortic position in our Pulse Duplicator System for simultaneous assessment of valve performance and valve leakage data. During forward flow conditions, measured mean pressure gradients are between 4.5 and 20.2 mmHg (11.9 +/- 4.4 mmHg) for cardiac outputs between 3.6 and 5.3 /min (4.5 +/- 0.4 /min). Doppler-derived mean pressure gradients are between 2.0 and 170 mmHg (9.3 +/- 3.9 mmHg) for the same flow conditions. Effective Orifice area is 1.31 +/- 0.08 cm2 and the performance index is 0.74 +/- 0.04, using the actual geometric orifice area, and 0.38 +/- 0.02, using the tissue annulus diameter, for a cardiac output of 4.5 +/- 0.4 l/min. Regurgitation volumes are below 3 ml. There is a trend to an effect of valve orientation on hemodynamics.
本研究的目的是在我们的脉搏复制系统中,对位于主动脉位置的Omnicarbon 21毫米心脏瓣膜进行流体动力学和多普勒超声心动图联合测试,以同时评估瓣膜性能和瓣膜泄漏数据。在前向血流条件下,心输出量在3.6至5.3升/分钟(4.5±0.4升/分钟)时,测得的平均压力阶差在4.5至20.2毫米汞柱之间(11.9±4.4毫米汞柱)。在相同血流条件下,多普勒得出的平均压力阶差在2.0至17.0毫米汞柱之间(9.3±3.9毫米汞柱)。对于心输出量为4.5±0.4升/分钟的情况,有效瓣口面积为1.31±0.08平方厘米,使用实际几何瓣口面积时性能指数为0.74±0.04,使用组织瓣环直径时性能指数为0.38±0.02。反流容积低于3毫升。瓣膜方向对血流动力学有影响的趋势。