Suppr超能文献

电子径向阵列超声内镜的初步经验:与机械扇形扫描超声内镜在人体中的随机对照研究。

Initial experience with an electronic radial array echoendoscope: randomized comparison with a mechanical sector scanning echoendoscope in humans.

作者信息

Anderson Michelle A, Scheiman James M

机构信息

Division of Gastroenterology, Department of Internal Medicine, Endoscopic Ultrasound Program, University of Michigan Health System, Ann Arbor, Michigan 48109, USA.

出版信息

Gastrointest Endosc. 2002 Oct;56(4):573-7. doi: 10.1067/mge.2002.127761.

Abstract

BACKGROUND

An electronic radial array echoendoscope has been developed that uses the same US processor as the linear array instruments made by the same company, thereby eliminating the need for two processors. The performance of this prototype instrument was compared with that of a mechanical radial echoendoscope in patients.

METHODS

Fourteen patients underwent sequential examinations with both echoendoscopes in random order. Predefined criteria were used to evaluate the performance of each instrument.

RESULTS

The quality of the electronic radial image was rated as "good" or "superior" to that of the mechanical radial echoendoscope in 12 patients. In 2 patients, imaging with the prototype was inferior because of positioning and/or operator inexperience. The forward-viewing optics of the prototype enhanced intubation and instrument advancement. Limitations included excessive stiffness of the tip of the instrument that resulted in occasional slippage from the duodenum. The prototype provided an adequate diagnosis in 12 patients and was superior to the mechanical radial endoscope in terms of diagnosis in 2 cases.

CONCLUSION

The prototype electronic radial echoendoscope provides US images of similar quality and is equivalent in terms of clinical utility to the mechanical radial echoendoscope.

摘要

背景

已研发出一种电子径向阵列超声内镜,它使用与同一家公司生产的线性阵列仪器相同的超声处理器,从而无需两台处理器。在患者中对该原型仪器的性能与机械径向超声内镜的性能进行了比较。

方法

14名患者按随机顺序先后接受了两种超声内镜的检查。使用预定义标准评估每种仪器的性能。

结果

12名患者的电子径向图像质量被评为“良好”或优于机械径向超声内镜。在2名患者中,由于定位和/或操作者经验不足,原型仪器的成像效果较差。原型仪器的前视光学装置便于插管和器械推进。局限性包括仪器尖端过硬,导致偶尔从十二指肠滑落。该原型仪器对12名患者做出了充分诊断,在2例诊断方面优于机械径向内镜。

结论

原型电子径向超声内镜提供了质量相似的超声图像,在临床实用性方面与机械径向超声内镜相当。