Tsai Chieh-Chih, Kau Hui-Chuan, Kao Shu-Ching, Hsu Wen-Ming
Department of Ophthalmology, Taipei Veterans General Hospital, Taiwan, ROC.
Ophthalmic Surg Lasers. 2002 Sep-Oct;33(5):430-2.
We present 2 patients with accidental corneal burns inflicted during clear cornea cataract surgery using radiofrequency diathermy for anterior capsulotomy. During the capsulotomy procedure, the anterior capsule was not opened and only a small air bubble appeared when the energy was turned on. Meanwhile the area of the corneal tunnel that contacted the shoulder of the diathermy tip became opaque and shrank with a small defect in the anterior corneal lip. Postoperatively, focal corneal shrinkage with iris incarceration into the defect of the corneal tunnel and resultant peaking pupil were noted in the 2 patients. The corneal burns resulted in temporary irregular astigmatism with mild visual acuity impairment. To our knowledge, this is the first report of a corneal burn as a complication of radiofrequency diathermy for anterior capsulotomy in clear cornea cataract surgery.
我们报告2例在透明角膜白内障手术中使用射频透热法进行前囊切开术时意外发生角膜烧伤的患者。在前囊切开术过程中,前囊未打开,打开能量时仅出现一个小气泡。与此同时,与透热尖端肩部接触的角膜隧道区域变得不透明并收缩,角膜前唇出现一个小缺损。术后,这2例患者均出现局限性角膜收缩,虹膜嵌顿于角膜隧道缺损处,导致瞳孔尖峰形成。角膜烧伤导致暂时性不规则散光,并伴有轻度视力损害。据我们所知,这是关于透明角膜白内障手术中射频透热法进行前囊切开术导致角膜烧伤并发症的首例报告。