Spaepen M S, Bodman H A, Kundsin R B, Berryman J R, Fencl V
Health Lab Sci. 1975 Oct;12(4):316-20.
Out of 353 heated nebulizers in actual use which were cultured in the Peter Bent Brigham Hospital, 118 (33%) were found to be contaminated. Gram-negative organisms predominated with rare micrococci and fungi. A slow-growing Vibrio was also recovered from 4 nebulizers. Heating of nebulizers above 46 C resulted in significant reduction of contamination. Types of heated nebulizers also figured significantly in the rate of contamination found. Carefully considered recommendations for the care and use of respiratory therapy equipment must be instituted and enforced. Techniques for terminal disinfection of equipment, followed by sterilization should result in the issue of sterile equipment for each patient. Rules for maintenance of equipment at the bedside are also needed.
在彼得·本特·布里格姆医院对实际使用的353台加热雾化器进行培养后发现,有118台(33%)受到污染。革兰氏阴性菌占主导,伴有罕见的微球菌和真菌。从4台雾化器中还分离出一种生长缓慢的弧菌。将雾化器加热到46摄氏度以上可显著减少污染。所发现的污染率还与加热雾化器的类型密切相关。必须制定并执行关于呼吸治疗设备护理和使用的审慎建议。设备的终末消毒技术,随后进行灭菌处理,应能确保为每位患者提供无菌设备。同时也需要制定床边设备维护规则。