Antoch Gerald, Freudenberg Lutz S, Egelhof Thomas, Stattaus Jörg, Jentzen Walter, Debatin Jörg F, Bockisch Andreas
Department of Diagnostic and Interventional Radiology, University Hospital Essen, Essen, Germany.
J Nucl Med. 2002 Oct;43(10):1339-42.
This study was performed to evaluate a possible artifact related to the administration of intravascular contrast agent in dual-modality PET/CT imaging.
Thirty oncology patients underwent whole-body PET/CT. CT images, which were collected in the presence of intravenous and oral iodinated contrast agent, were used for PET attenuation correction. PET images were assessed for the artifact, defined as a region of high count rate on attenuation-corrected images in accurate coregistration with a contrast-enhanced blood vessel. Intravascular enhancement of thoracic veins was quantified by application of regions of interest, and quantities in patients with the artifact (group 1) and without the artifact (group 2) were correlated. Body surface area was calculated for all patients.
The contrast-induced PET artifact was present in 4 (13%) of 30 patients. Mean density differences in intravascular enhancement were highly significant (P < 0.001) in a comparison of group 1 (2,262 +/- 304 Hounsfield units [HU]) and group 2 (1,058 +/- 209 HU). Body surface area was significantly lower (P = 0.035) in the patients of group 1 (1.67 +/- 0.11 m(2)) than in the patients of group 2 (2.01 +/- 0.18 m(2)).
Contrast-enhanced dual-modality PET/CT examinations may result in a PET artifact that is due to the transient bolus passage of undiluted intravenous contrast agent.
本研究旨在评估在双模态PET/CT成像中血管内造影剂给药可能产生的伪影。
30例肿瘤患者接受全身PET/CT检查。在静脉和口服碘化造影剂存在的情况下采集的CT图像用于PET衰减校正。对PET图像进行伪影评估,伪影定义为在与增强血管精确配准的衰减校正图像上的高计数率区域。通过应用感兴趣区域对胸段静脉的血管内强化进行量化,并将有伪影的患者(第1组)和无伪影的患者(第2组)的量化结果进行相关性分析。计算所有患者的体表面积。
30例患者中有4例(13%)出现了对比剂诱导的PET伪影。在第1组(2262±304亨氏单位[HU])和第2组(1058±209 HU)的比较中,血管内强化的平均密度差异非常显著(P<0.001)。第1组患者(1.67±0.11 m²)的体表面积显著低于第2组患者(2.01±0.18 m²)(P = 0.035)。
对比增强双模态PET/CT检查可能会导致PET伪影,这是由于未稀释的静脉造影剂的短暂团注通过所致。