在口腔癌和头颈癌的PET/CT研究中,减少牙种植体的CT伪影是否会影响PET图像解读?

Does reducing CT artifacts from dental implants influence the PET interpretation in PET/CT studies of oral cancer and head and neck cancer?

作者信息

Nahmias Claude, Lemmens Catherine, Faul David, Carlson Eric, Long Misty, Blodgett Todd, Nuyts Johan, Townsend David

机构信息

Molecular Imaging and Translational Research Program, University of Tennessee, Knoxville, Tennessee 37920-6999, USA.

出版信息

J Nucl Med. 2008 Jul;49(7):1047-52. doi: 10.2967/jnumed.107.049858. Epub 2008 Jun 13.

Abstract

UNLABELLED

In patients with oral head and neck cancer, the presence of metallic dental implants produces streak artifacts in the CT images. These artifacts negate the utility of CT for the spatial localization of PET findings and may propagate through the CT-based attenuation correction into the PET images. In this study, we evaluated the efficacy of an algorithm that reduces metallic artifacts in CT images and the impact of this approach on the quantification of PET images.

METHODS

Fifty-one patients with and 9 without dental implants underwent a PET/CT study. CT images through the patient's dental implants were reconstructed using both standard CT reconstruction and an algorithm that reduces metallic artifacts. Attenuation correction factors were calculated from both sets of CT images and applied to the PET data. The CT images were evaluated for any reduction of the artifacts. The PET images were assessed for any quantitative change introduced by metallic artifact reduction.

RESULTS

For each reconstruction, 2 regions of interest were defined in areas where the standard CT reconstruction overestimated the Hounsfield units (HU), 2 were defined in underestimated areas, and 1 was defined in a region unaffected by the artifacts. The 5 regions of interest were transferred to the other 3 reconstructions. Mean HU or mean Bq/cm(3) were obtained for all regions. In the CT reconstructions, metallic artifact reduction decreased the overestimated HUs by approximately 60% and increased the underestimated HUs by approximately 90%. There was no change in quantification in the PET images between the 2 algorithms (Spearman coefficient of rank correlation, 0.99). Although the distribution of attenuation (HU) changed considerably in the CT images, the distribution of activity did not change in the PET images.

CONCLUSION

Our study demonstrated that the algorithm can enhance the structural and spatial content of CT images in the presence of metallic artifacts. The CT artifacts do not propagate through the CT-based attenuation correction into the PET images, confirming the robustness of CT-based attenuation correction in the presence of metallic artifacts. The study also demonstrated that considerable changes in CT images do not change the PET images.

摘要

未标注

在口腔头颈癌患者中,金属牙种植体的存在会在CT图像中产生条纹伪影。这些伪影使CT在PET检查结果空间定位方面的效用丧失,并且可能通过基于CT的衰减校正传播到PET图像中。在本研究中,我们评估了一种减少CT图像中金属伪影的算法的效果以及该方法对PET图像定量分析的影响。

方法

51例有牙种植体和9例无牙种植体的患者接受了PET/CT检查。通过患者牙种植体的CT图像采用标准CT重建和一种减少金属伪影的算法进行重建。从两组CT图像计算衰减校正因子并应用于PET数据。评估CT图像中伪影的减少情况。评估PET图像中因减少金属伪影而引入的任何定量变化。

结果

对于每次重建,在标准CT重建高估亨氏单位(HU)的区域定义2个感兴趣区,在低估区域定义2个,在不受伪影影响的区域定义1个。将这5个感兴趣区转移到其他3次重建中。获取所有区域的平均HU或平均Bq/cm³。在CT重建中,减少金属伪影使高估的HU降低约60%,使低估的HU增加约90%。两种算法之间PET图像的定量分析没有变化(斯皮尔曼等级相关系数为0.99)。尽管CT图像中衰减(HU)的分布有很大变化,但PET图像中活性分布没有变化。

结论

我们的研究表明,该算法在存在金属伪影的情况下可增强CT图像的结构和空间内容。CT伪影不会通过基于CT的衰减校正传播到PET图像中,证实了在存在金属伪影时基于CT的衰减校正的稳健性。该研究还表明,CT图像中的显著变化不会改变PET图像。