Abstract

The power density at the retina provided by presently available laser photocoagulators is subject to variation. Three factors contribute to making the retinal image of the laser beam substantially different from the spot size indicated on the selector of the photocoagulator. These are the operator's variable accommodation, the astigmatism introduced by the contact lens, and the depth of focus of the observation system. The assumption that the same setting on the instrument regularly produces the same power density at the retina will lead to errors in comparing lesions. The safest procedure is to start from the threshold in every location.

摘要

目前可用的激光光凝器在视网膜上提供的功率密度存在变化。有三个因素导致激光束在视网膜上的图像与光凝器选择器上指示的光斑大小有很大不同。这些因素包括操作者可变的调节能力、接触镜引入的散光以及观察系统的焦深。认为仪器上相同的设置会在视网膜上定期产生相同功率密度的假设,会导致在比较病变时出现误差。最安全的做法是在每个位置都从阈值开始。